Model Number PM2272
Device Problems Failure to Interrogate (1332); No Pacing (3268)
Patient Problems Arrhythmia (1721); Fatigue (1849)
Event Date 12/16/2022
Event Type  Injury  
Event Description
It was reported that the patient experienced bradycardia and fatigue.The device could not be interrogated and was not pacing.The device was explanted and replaced to resolve the event.The patient was stable following the procedure and there were no adverse consequences.
 
Manufacturer Narrative
The reported event of inability to interrogate was confirmed.The reported event of inability to pace was not confirmed.The device was received with intermittent telemetry communication and normal output.Visual inspection of the header attachment area detected an anomaly between the epoxy header and titanium case.The device was cut open to enable further testing and the battery was found at normal levels.Feedthrough leak test was performed, indicating a device hermeticity breach.This is consistent with feedthrough damage as a result of fluid intrusion between the header and case, and subsequent fluid ingress to internal electronics.Hybrid circuitry was tested, and the results indicated normal current drain.A manufacturing process anomaly may have occurred, which resulted in the header bonding anomaly.The device is included in the zenex, assurity, endurity laser adhesion preparation advisory issued by abbott on 20 july 2022 for a subset of devices distributed and implanted outside of the united states.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ASSURITY MRI
Type of Device
IMPLANTABLE PACEMAKER PULSE GENERATOR
Manufacturer (Section D)
ABBOTT
15900 valley view court
sylmar CA 91342
Manufacturer (Section G)
ABBOTT
15900 valley view court
sylmar CA 91342
Manufacturer Contact
richard williamson
15900 valley view court
sylmar, CA 91342
MDR Report Key16039598
MDR Text Key306061013
Report Number2017865-2022-50546
Device Sequence Number1
Product Code LWP
Combination Product (y/n)N
Reporter Country CodeDA
PMA/PMN Number
P140033
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 03/21/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/22/2022
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date08/31/2017
Device Model NumberPM2272
Device Catalogue NumberPM2272
Device Lot NumberA000023386
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received02/22/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured02/26/2016
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
-
-