Model Number N/A
Medical Device Problem Code Material Deformation (2976)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 12/04/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The manufacturer has received the sample and is pending evaluation.Results are expected soon.A lot history review (lhr) of regt4077 showed one other similar product complaint(s) from this lot number.The complaints for this lot number have been reported from the same facility.
 
Event or Problem Description
It was reported by the customer "per vat team member "two 3fs picc introducers are long and flimsy.One has a white ridge and the inserter was not able to get it through the patient's skin.The second one got a burr on the end trying to insert in the patient's skin and a few notches in the outside channel." this report addresses the inserter with the burr at the end.
 
Event or Problem Description
It was reported by the customer "per vat team member "two 3fs picc introducers are long and flimsy.One has a white ridge and the inserter was not able to get it through the patient's skin.The second one got a burr on the end trying to insert in the patient's skin and a few notches in the outside channel." this report addresses the inserter with the burr at the end.
 
Additional Manufacturer Narrative
H11: section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.The following were reviewed as part of this investigation: patient severity, complaint and batch history, applicable previous investigation(s), applicable manufacturing records, photo analysis and applicable fmea documents.Based on a review of this information, the following was concluded: the complaint of difficult insertion of the microintroducers was inconclusive due to the sample condition.A photograph of the two implicated 3.5fr ptfe microintroducers was returned for evaluation.Both samples were found within plastic bags.One of the samples consisted of the dilator within the sheath while the other sample was the sheath without the inlaid dilator.The sheath had not been and peeled on either sample.The distal end of the dilator and sheaths could not be clearly visualized in the returned photograph.No close-up photographs were provided of the reported ridges or burrs seen on the devices.The submitted photograph did not contain distinguishing features that could aid in confirmation of the complaint nor identification of a specific root cause.Therefore, this complaint will be recorded as inconclusive at this time.Possible contributing factors could include too small of an incision, a steep insertion angle, a poor transition of the introducer sheath to the dilator, or damage to the dilator or sheath prior to use.Sheath damage can be minimized by simultaneously advancing the sheath and dilator as a single unit using a rotational motion.If necessary, a small incision may be made adjacent to the guidewire to facilitate insertion of the sheath and dilator.H3 other text : evaluation findings are in section h.11.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DOT PROV 3F SL MAX 3CG
Common Device Name
CATHETER,INTRAVASCULAR,THERAPEUTIC,LONG-TERM GREATER THAN 30 DAYS
Manufacturer (Section D)
C.R. BARD, INC. (BASD) -3006260740
605 north 5600 west
salt lake city 84116
Manufacturer (Section G)
BARD REYNOSA S.A. DE C.V. -9617592
blvd. montebello #1
parque industrial colonial
reynosa, tamaulipas 88780
MX   88780
Manufacturer Contact
kayla olsen
605 north 5600 west
salt lake city 84116
8015950700
MDR Report Key16039723
Report Number3006260740-2022-05828
Device Sequence Number20143014
Product Code LJS
UDI-Device Identifier00801741130342
UDI-Public(01)00801741130342
Combination Product (Y/N)N
Initial Reporter StateNC
Initial Reporter CountryUS
PMA/510(K) Number
K162443
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Other,User Facility
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 02/07/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date10/31/2023
Device Model NumberN/A
Device Catalogue NumberS1173108D
Device Lot NumberREGT4077
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/06/2022
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 12/04/2022
Supplement Date Received by Manufacturer02/03/2023
Initial Report FDA Received Date12/22/2022
Supplement Report FDA Received Date02/11/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured06/01/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient SexUnknown
-
-