• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SISTEMAS MEDICOS ALARIS, S.A. DE C.V. BD SMARTSITE¿ EXTENSION SET, SMALLBORE SLIDE CLAMP; INTRAVASCULAR ADMINISTRATION SET

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SISTEMAS MEDICOS ALARIS, S.A. DE C.V. BD SMARTSITE¿ EXTENSION SET, SMALLBORE SLIDE CLAMP; INTRAVASCULAR ADMINISTRATION SET Back to Search Results
Catalog Number 20043E
Medical Device Problem Code Leak/Splash (1354)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 12/05/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
A device evaluation is anticipated but has not yet begun.Upon completion of the investigation, a supplemental report will be filed.
 
Event or Problem Description
It was reported that the bd smartsite¿ extension set, smallbore slide clamp was leaking.The following information was provided by the initial reporter: the nurse noticed the first one leaking, switched to another one that was still leaking, and finally got a third one out that didn¿t leak.It was connected to a clave that was connected to a normal saline flush.It was leaking the normal saline, and was discovered when priming the tubing prior to patient use.
 
Additional Manufacturer Narrative
The following fields were updated due to additional information: d10: device available for eval yes.D10: returned to manufacturer on: 19-dec-2022.H6: investigation summary one sample was received and tested by our quality team.The set was visually inspected and there were no issues discovered.Saline was then infused through the device and no leak was detected.Finally the end of the set was clamped and saline was injected into set to try to force a leak but still there was nothing realized.The complaint of leakage could not be verified and no root cause can be established.A device history record review for model 20043e lot number 22095288 was performed.There were no quality notifications issued for the failure mode reported by the customer during the build of this set.
 
Event or Problem Description
It was reported that the bd smartsite¿ extension set, smallbore slide clamp was leaking.The following information was provided by the initial reporter: the nurse noticed the first one leaking, switched to another one that was still leaking, and finally got a third one out that didn¿t leak.It was connected to a clave that was connected to a normal saline flush.It was leaking the normal saline, and was discovered when priming the tubing prior to patient use.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BD SMARTSITE¿ EXTENSION SET, SMALLBORE SLIDE CLAMP
Common Device Name
INTRAVASCULAR ADMINISTRATION SET
Manufacturer (Section D)
SISTEMAS MEDICOS ALARIS, S.A. DE C.V.
blvd. insurgentes no. 20351
parque industrial el florido
tijuana
Manufacturer (Section G)
SISTEMAS MEDICOS ALARIS, S.A. DE C.V.
blvd. insurgentes no. 20351
parque industrial el florido
tijuana
Manufacturer Contact
phillip emmert
9450 south state street
sandy, UT 84070
8015296192
MDR Report Key16039973
Report Number9616066-2022-02051
Device Sequence Number14683562
Product Code FPA
UDI-Device Identifier50885403234871
UDI-Public10885403234873
Combination Product (Y/N)N
Initial Reporter StateOH
Initial Reporter CountryUS
PMA/510(K) Number
K960280
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Other,User Facility
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 02/08/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Other
Device Expiration Date09/14/2025
Device Catalogue Number20043E
Device Lot Number22095288
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 12/05/2022
Supplement Date Received by Manufacturer02/14/2023
Initial Report FDA Received Date12/22/2022
Supplement Report FDA Received Date02/23/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured09/05/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-