• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: AVANOS MEDICAL INC. CORFLO NG/NI FT W/ STYLET & ENFIT; DH CPK NG TUBES

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

AVANOS MEDICAL INC. CORFLO NG/NI FT W/ STYLET & ENFIT; DH CPK NG TUBES Back to Search Results
Model Number 40-9438
Medical Device Problem Code Material Integrity Problem (2978)
Health Effect - Clinical Code Feeding Problem (1850)
Date of Event 11/08/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The sample is reported to be available but has not yet been received by the manufacturer.A review of the device history record is in-progress.All information reasonably known as of 22 dec 2022 has been included in this health authority report.Should additional information be obtained, a follow-up health authority report will be provided.The information provided by avanos medical inc.Represents all of the known information at this time.Avanos medical inc.Has no independent knowledge of the event reported but is relaying the information that was provided by the user facility where the incident occurred.This product incident is documented in the avanos medical complaint database and identified as complaint (b)(4).This information is submitted pursuant to 21cfr803, in compliance with the medical device reporting requirement and should not be considered to be an admission that an avanos medical inc.Product is defective or caused serious injury.
 
Event or Problem Description
Fda medwatch / fda user facility report # mw (b)(4).Reported, the infant was intubated.A chest x-ray was used revealed dilatation/ballooning at the cervical portion of the corflo nasogastric/nasointestinal (ng) feeding tube.A second x-ray was completed which showed focal discontinuity of the corflo tube at the level of the thoracic inlet.The ng tube was removed in its entirety showing an aneurysmal section with lumen exposed.It was additionally reported, the ng tube was being used for tube feed; however, ¿the device did not do what it was supposed to do.¿ there was no reported injury.
 
Event or Problem Description
Additional information received 22dec2022 reported, the tube was placed on (b)(6) 2022 and was removed on (b)(6) 2022; the tube integrity was noticed on an x-ray.The syringe size used for small children was no more than 10mls; all of the meds were liquid.
 
Additional Manufacturer Narrative
The device history record for lot 30208011was reviewed and the product was produced according to product specifications.All information reasonably known as of 17 jan 2023 has been included in this health authority report.Should additional information be obtained, a follow-up health authority report will be provided.The information provided by avanos medical inc.Represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to avanos medical inc.Avanos medical inc.Has no independent knowledge of the event reported but is relaying the information that was provided by the user facility where the incident occurred.This product incident is documented in the avanos medical complaint database and identified as complaint (b)(4).This information is submitted pursuant to 21cfr803, in compliance with the medical device reporting requirement and should not be considered to be an admission that an avanos medical inc.Product is defective or caused serious injury.
 
Additional Manufacturer Narrative
The actual sample from the reported event was returned for evaluation.Visual examination of the device revealed the tubing had signs of damage and discoloration.Testing of the sample: when the tubing was manually pressed back together, thin layers of the material were noticed on the ballooned portion of the tube and both breaking points exhibited unknown build-up residues.The reported event could be confirmed as reported: the tube expanded to form a balloon shape which burst and caused a separation of the tube.The root cause is undetermined.All information reasonably known as of 08 feb 2023 has been included in this health authority report.Should additional information be obtained, a follow-up health authority report will be provided.The information provided by avanos medical inc.Represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to avanos medical inc.Avanos medical inc.Has no independent knowledge of the event reported but is relaying the information that was provided by the user facility where the incident occurred.This product incident is documented in the avanos medical complaint database and identified as complaint (b)(4).This information is submitted pursuant to 21cfr803, in compliance with the medical device reporting requirement and should not be considered to be an admission that an avanos medical inc.Product is defective or caused serious injury.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CORFLO NG/NI FT W/ STYLET & ENFIT
Common Device Name
DH CPK NG TUBES
Manufacturer (Section D)
AVANOS MEDICAL INC.
5405 windward parkway
alpharetta GA 30004
Manufacturer (Section G)
AVENT SA DE R.L. DE C.V. (AVENT 1)
circuito industial no.40
colonia obrera
nogales, cp
Manufacturer Contact
lisa clark
5405 windward parkway
alpharetta, GA 30004
4704485444
MDR Report Key16040068
Report Number9611594-2022-00159
Device Sequence Number5306734
Product Code KNT
UDI-Device Identifier00350770460505
UDI-Public00350770460505
Combination Product (Y/N)N
Initial Reporter StateAZ
Initial Reporter CountryUS
PMA/510(K) Number
K821906
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Other,Health Professional,User Facility
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup,Followup
Report Date (Section B) 02/08/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number40-9438
Device Lot Number30208011
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 11/29/2022
Supplement Date Received by Manufacturer12/22/2022
01/12/2023
Initial Report FDA Received Date12/22/2022
Supplement Report FDA Received Date01/17/2023
02/08/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured07/05/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Patient Age6 MO
Patient SexMale
Patient Weight9 KG
-
-