| Model Number |
101/870/090CZ |
| Medical Device Problem Codes |
Disconnection (1171); Gas/Air Leak (2946)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
09/01/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that during the use of the product, leakage of air from it was observed.The customer checked the product and the pilot balloon was found detached.No patient injury reported.
|
| |
|
Additional Manufacturer Narrative
|
|
Device serial number/lot number is unknown, no finished good lot number provided.Returned device have laser printed lot on its flange, unable to verify.Device manufacturing date and device expiration date could not be determined a product sample was received for evaluation.Visual and functional testing were performed.The inflation line was detached from pilot balloon.The root cause of the reported issue was found to be an inflation system leak due to supplied item manufacturing fault.Manufacturing device history record review was not performed because the results of the complaint investigation is a supplier design related problem - currently investigated within capa and trend analysis.
|
| |
|