• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DAVIS & GECK CARIBE LTD SURGIDAC; ENDOSCOPIC TISSUE APPROXIMATION DEVICE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DAVIS & GECK CARIBE LTD SURGIDAC; ENDOSCOPIC TISSUE APPROXIMATION DEVICE Back to Search Results
Model Number 170044
Medical Device Problem Codes Break (1069); Detachment of Device or Device Component (2907); Human-Device Interface Problem (2949)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 12/07/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Concomitant medical products: 173016 - 173016 endo stitch instrument, lot#j2g0204ey 173016 - 173016 endo stitch instrument, lot#j2g0204ey 173016 - 173016 endo stitch instrument, lot#j2g0204ey 170044 -170044 endo stitch tri 2-0 grn 120 s'dac, lot# j2b2352y 170044 -170044 endo stitch tri 2-0 grn 120 s'dac, lot# j2b2352y 170044 - 170044 endo stitch tri 2-0 grn 120 s'dac, lot# unknown 170044 - 170044 endo stitch tri 2-0 grn 120 s'dac, lot# unknown.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
According to the reporter, on a laparoscopic roux-en-y gastric bypass and hiatal hernia repair, during the creation of side-to-side anastomosis, the devices had unloading difficulties.The needles broke, disengaged from the handle, and fell into the patient¿s cavity.The needles were retrieved with a laparoscopic grasper.A single-use suturing device was used instead.There was no patient injury.
 
Additional Manufacturer Narrative
Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.
 
Event or Problem Description
According to the reporter, on a laparoscopic roux-en-y gastric bypass and hiatal hernia repair, during the creation of side-to-side anastomosis, the devices had unloading difficulties.Three needles broke, disengaged from the handle, and fell into the patient¿s cavity.All three needles were retrieved with a laparoscopic grasper.A single-use suturing device was used instead.There was no patient injury.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SURGIDAC
Common Device Name
ENDOSCOPIC TISSUE APPROXIMATION DEVICE
Manufacturer (Section D)
DAVIS & GECK CARIBE LTD
zona franca de san isidro
santo domingo 0101
DR  0101
Manufacturer (Section G)
DAVIS & GECK CARIBE LTD
zona franca de san isidro
santo domingo 0101
DR   0101
Manufacturer Contact
justin ellis
8200 coral sea st ne
mounds view, MN 55112
7635265677
MDR Report Key16040382
Report Number9612501-2022-02239
Device Sequence Number6437620
Product Code OCW
UDI-Device Identifier20884521127019
UDI-Public20884521127019
Combination Product (Y/N)N
Initial Reporter StateIL
Initial Reporter CountryUS
PMA/510(K) Number
K934738
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 01/20/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number170044
Device Catalogue Number170044
Device Lot NumberJ2B2352Y
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 12/07/2022
Supplement Date Received by Manufacturer01/05/2023
Initial Report FDA Received Date12/22/2022
Supplement Report FDA Received Date01/20/2023
Date Device Manufactured02/22/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
SEE H10.
Patient Age56 YR
Patient SexFemale
-
-