INTEGRA LIFESCIENCES CORPORATION OH/USA ULTRA 360 PATIENT POSITIONING SYSTEM; N/A Back to Search Results
Model Number A2009
Medical Device Problem Code Device Slipped (1584)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The ultra 360 patient positioning system (a2009) was returned for evaluation: failure analysis - complaint confirmed via inspection of the unit.The upper and lower handles were adjusted to bring them back into tolerance from not holding properly.Both shock cushions, 1/4in pin and adjustment wrench were replaced from being worn.Unit received with std.Adaptor and not the tri-star adaptor.General maintenance and cleaning performed.All worn components replaced with new parts.Handles adjusted.Root cause - probable root cause is routine use and wear of the unit.This unit is beyond integra¿s 7 years recommended life cycle (manufactured 2008).No further investigation is required based on the acceptability of risk and no adverse trends were identified.This will be monitored and trended going forward.At present, we consider this complaint to be closed.
 
Event or Problem Description
A facility reported that the ultra 360 patient positioning system (a2009) slips when locked (extrusion of shock cushion on upper locking arm).No information has been provided on patient involvement, injury, or surgical delay.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ULTRA 360 PATIENT POSITIONING SYSTEM
Common Device Name
N/A
Manufacturer (Section D)
INTEGRA LIFESCIENCES CORPORATION OH/USA
4900 charlemar drive
4900 charlemar drive
cincinnati OH
Manufacturer (Section G)
INTEGRA LIFESCIENCES CORPORATION OH/USA
4900 charlemar drive
cincinnati OH
Manufacturer Contact
vivian nelson
1100 campus drive
princeton, NJ 
6099362319
MDR Report Key16040393
Report Number3004608878-2022-00287
Device Sequence Number20399716
Product Code FWZ
UDI-Device Identifier10381780267850
UDI-Public10381780267850
Combination Product (Y/N)N
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source User Facility,Company Representative
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 12/22/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberA2009
Device Catalogue NumberA2009
Device Lot Number081
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/28/2022
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 12/05/2022
Initial Report FDA Received Date12/22/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1