| Model Number |
1053958 |
| Medical Device Problem Codes |
Calibration Problem (2890); Device-Device Incompatibility (2919); Compatibility Problem (2960)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
12/12/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The manufacturer's investigation is on-going and further information will be provided once the investigation has been completed.
|
| |
|
Event or Problem Description
|
|
The customer reported accuracy deviations with leksell vantage during surgery with a systematic error of about 1.6 mm.The customer was using the non-compatible microdrive from the manufacturer alpha-omega.
|
| |
|
Additional Manufacturer Narrative
|
|
D8 updated.H6 updated.H10 updated: the issue reported by the customer was that the leksell vantage stereotactic system had a target shift of more than 2mm.This site is using a microdrive supplied by a third party (alpha omega) in conjunction with the leksell vantage stereotactic system.The leksell vantage stereotactic system has been investigated regarding mechanical accuracy and it is well within specification.Elekta instrument ab have also investigated the used microdrive (alpha omega) together with the leksell vantage stereotactic system and the investigation concluded that microdrive (alpha omega) is incorrectly interfacing the leksell vantage stereotactic system in a way that it may cause a significant shift both anterior and lateral (accuracy error).This shift is seen for the system tested only and is judged to be highly dependent on the combination of microdrive (alpha omega) measurements and the stop holder measurements.The result was presented to the user who confirms that the direction is the same as they have seen during their procedures as well as during phantom tests.Tests were also repeated with the leksell vantage stereotactic system using elekta microdrive which show much better accuracy (within specification) than using the alpha omega microdrive due to correct interfacing to the system.The root cause could therefore be concluded to be caused by incompatibility by a third party product.Alpha omega is claiming compatibility with all elekta instrument ab's stereotactic systems on the market.Elekta instrument ab has on several occasions informed alpha omega that interfacing towards only the stop holder is incorrect and will cause accuracy issues.Elekta instrument ab recommended alpha omega to review their interfacing design and stated that alpha omega cannot claim compatibility with elekta instrument ab's stereotactic systems.Leksell vantage stereotactic system inhibited safely and there have been no patient injuries.Please note that there are two cases reported from the same user using the same system since it was used twice where a discrepancy was seen (elekta ref.#(b)(4) - mfr no 9612186-2022-00004).
|
| |
|
Search Alerts/Recalls
|