| Model Number |
SA60AT |
| Medical Device Problem Codes |
Defective Device (2588); Appropriate Device Problem Term/Code Not Available (3191)
|
| Health Effect - Clinical Codes |
Insufficient Information (4580); No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
11/28/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
A materials manager reported that during a cataract extraction with an intraocular lens (iol) implant procedure, the lens would not load and kept moving around.Patient contact was observed and procedure was completed with another lens.Additional information was requested, but further no information was available.
|
| |
|
Additional Manufacturer Narrative
|
|
A sample device was not returned for analysis.Product history records were reviewed and documentation indicated the product met release criteria.Root cause has not been identified.There have been no other complaints reported in the lot number.The manufacturer internal reference number is: (b)(4).
|
| |
|
Additional Manufacturer Narrative
|
|
Additional information was provided in a.1, a.2, a.3, b.5, d.10, e.1, e.4, and h.6 the manufacturer internal reference number is: (b)(4).
|
| |
|
Event or Problem Description
|
|
Additional information was received stating there was no patient harm.It was reported that patient's issues were resolved.
|
| |
|
Additional Manufacturer Narrative
|
|
A non-qualified cartridge was indicated with a qualified cartridge and viscoelastic.The company, 33.0 diopter lens is only qualified for use in the company cartridge.The root cause for the reported loading issue may to be related to a failure to follow the ifu (instructions for use).The ifu (instructions for use) instructs that a company qualified delivery system and viscoelastic combination should be used.The use of an unqualified combination may cause damage to the lens and potential complications during the implantation process.Company foldable intraocular lenses are qualified for use with a company qualified delivery system (handpiece and cartridge) and ophthalmic viscosurgical device (ovd) combination.The manufacturer internal reference number is: (b)(4).
|
| |
|