• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTRONIC SOFAMOR DANEK USA, INC T2¿ SPINAL SYSTEM; SPINAL VERTEBRAL BODY REPLACEMENT DEVICE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC SOFAMOR DANEK USA, INC T2¿ SPINAL SYSTEM; SPINAL VERTEBRAL BODY REPLACEMENT DEVICE Back to Search Results
Model Number 4360113
Medical Device Problem Codes Material Twisted/Bent (2981); Mechanics Altered (2984)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 11/30/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
Information was received from a healthcare professional via manufacturer representative regarding an inserter used in t12 vertebral colum resection/t2-ilium psf therapy for scoliosis.It was reported that the inserter would not release from the corpectomy cage while inside the patient.The doctor struggled to remove the inserter and twisted to remove it, causing the distal tip to bend.No patient symptoms were reported.No further complications were reported/anticipated.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
T2¿ SPINAL SYSTEM
Common Device Name
SPINAL VERTEBRAL BODY REPLACEMENT DEVICE
Manufacturer (Section D)
MEDTRONIC SOFAMOR DANEK USA, INC
4340 swinea rd
memphis TN 38118
Manufacturer (Section G)
MEDTRONIC SOFAMOR DANEK USA, INC
4340 swinea rd
memphis TN 38118
Manufacturer Contact
glen belmer
1800 pyramid place
memphis, TN 38132
6122713209
MDR Report Key16041839
Report Number1030489-2022-01126
Device Sequence Number6437402
Product Code MQP
UDI-Device Identifier00643169709447
UDI-Public00643169709447
Combination Product (Y/N)N
Initial Reporter StateIL
Initial Reporter CountryUS
PMA/510(K) Number
K173125
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 12/23/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number4360113
Device Catalogue Number4360113
Device Lot NumberRS19F004
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 11/30/2022
Initial Report FDA Received Date12/23/2022
Date Device Manufactured07/03/2019
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Patient Age68 YR
Patient SexFemale
-
-