| Model Number |
LI61AO |
| Medical Device Problem Codes |
Device Dislodged or Dislocated (2923); Material Integrity Problem (2978)
|
| Health Effect - Clinical Codes |
Corneal Scar (1793); Visual Impairment (2138); Eye Pain (4467)
|
| Date of Event |
11/14/2022
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
According to the reporter, device was discarded and is not available for evaluation.A device history record (dhr) review did not find any non-conformities or anomalies related to this event.The trend analysis, risk analysis and/or directions for use review were considered acceptable, with the product performing within anticipated rates.Based on the available information, user related factors such as loading or handling techniques) and/or procedural factors (such as lens and inserter interaction) may have caused or contributed to this event.
|
| |
|
Event or Problem Description
|
|
Reportedly, after implantation of an intraocular lens (iol) into the left eye it was discovered that iol haptics were bent.The surgeon attempted to straighten the haptics and during the process the iol flipped over and scratched the cornea.On post-op visit it was discovered that lens had shifted causing pain and visual problems for the patient.The lens was explanted 3 days post implant due to lens dislocation and extruded lens in the anterior chamber.A vitrectomy was performed, and the lens was replaced with another iol of the same model and diopter.
|
| |
|
Event or Problem Description
|
|
Additional information was received indicating the patient had a capsule rupture that occurred during cataract emulsification prior to the iol being placed.Therefore the lens was implanted in the sulcus.In the surgeons opinion the likely cause of this event is the kinked haptic in two locations leading to lens instability and subsequent dislocation in the anterior chamber.
|
| |
|
Additional Manufacturer Narrative
|
|
G4: added udi.
|
| |
|