• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEPUY SYNTHES PRODUCTS LLC 11.0 CM LONG ATTACHMENT; DRILL, SURGICAL, ENT (ELECTRIC OR PNEUMATIC)

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY SYNTHES PRODUCTS LLC 11.0 CM LONG ATTACHMENT; DRILL, SURGICAL, ENT (ELECTRIC OR PNEUMATIC) Back to Search Results
Model Number LONG
Medical Device Problem Codes Complete Loss of Power (4015); Excessive Heating (4030)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 01/01/2022
Type of Reportable Event Malfunction
Event or Problem Description
It was reported from india that during service and evaluation, it was determined that the attachment device was heating up.It was further determined that the device failed pretest for temperature.It was noted in the service order that the device was not working.This event did not occur during surgery.There was no patient involvement.There were no reports of injuries, medical intervention or prolonged hospitalization.The date of event was unknown but was known to have occurred in 2022.All available information has been disclosed.If additional information should become available, a supplemental medwatch will be submitted accordingly.
 
Additional Manufacturer Narrative
This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes, or its employees that the report constitutes an admission that the product, depuy synthes, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.Additional narrative: the actual device was returned for evaluation.The device was evaluated and it was determined that the reported condition was not confirmed.Therefore, the assignable root cause was not determined.However, during evaluation, it was determined that the device was heating up.It was further determined that the device failed pretest for temperature.The assignable root cause was traced to component failure due to normal wear.Udi: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
11.0 CM LONG ATTACHMENT
Common Device Name
DRILL, SURGICAL, ENT (ELECTRIC OR PNEUMATIC)
Manufacturer (Section D)
DEPUY SYNTHES PRODUCTS LLC
4500 riverside drive
palm beach gardens FL 33410
Manufacturer Contact
kate karberg
4500 riverside drive
palm beach gardens, FL 33410
3035526892
MDR Report Key16042717
Report Number1045834-2022-01762
Device Sequence Number17751706
Product Code HBC
UDI-Device Identifier00845384003571
UDI-Public00845384003571
Combination Product (Y/N)N
Initial Reporter CountryIN
PMA/510(K) Number
K011444
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,User Facility,Company Representative
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 12/23/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberLONG
Device Catalogue NumberLONG
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/16/2022
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 12/22/2022
Initial Report FDA Received Date12/23/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured01/22/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Patient SexUnknown
-
-