• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: CARDINAL HEALTH DOBBHOFF 10FR 43IN W STYLET EN; GASTROINTESTINAL TUBES WITH ENTERAL SPECIFIC CONNECTORS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

CARDINAL HEALTH DOBBHOFF 10FR 43IN W STYLET EN; GASTROINTESTINAL TUBES WITH ENTERAL SPECIFIC CONNECTORS Back to Search Results
Model Number 8884711006E
Medical Device Problem Code Detachment of Device or Device Component (2907)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 12/19/2022
Type of Reportable Event Malfunction
Event or Problem Description
The customer reported that the connector tip is detached from the feeding tube.The feeding tube is still in the patient.
 
Additional Manufacturer Narrative
An investigation is currently underway.Upon completion, the results will be forwarded.
 
Additional Manufacturer Narrative
The device history record (dhr) was reviewed, and no abnormal process conditions were present during the manufacturing of the product that could have led to the reported issue.The dhr review showed that all acceptance criteria inspections per established sampling levels were within acceptable limits during the production process.An investigation was carried out with the multifunctional team.All processes and controls were found to be followed correctly, including packaging and all inspections performed on the product.No abnormal conditions were found that could trigger the reported issue.One (1) photo was provided for evaluation.Through the photo, the reported issue was observed; the yport was detached.A corrective and preventive action has been initiated to address the reported issue.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DOBBHOFF 10FR 43IN W STYLET EN
Common Device Name
GASTROINTESTINAL TUBES WITH ENTERAL SPECIFIC CONNECTORS
Manufacturer (Section D)
CARDINAL HEALTH
777 west street
mansfield MA 02048
Manufacturer (Section G)
CARDINAL HEALTH
calle 9 sur no. 125 cuidad ind
tijuana 22500
MX   22500
Manufacturer Contact
jill saraiva
777 west street
mansfield, MA 02048
5086183640
MDR Report Key16042899
Report Number9612030-2022-03499
Device Sequence Number6238155
Product Code PIF
UDI-Device Identifier10884521582903
UDI-Public10884521582903
Combination Product (Y/N)N
Initial Reporter CountryCA
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,User Facility,Company Representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 01/23/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Model Number8884711006E
Device Catalogue Number8884711006E
Device Lot Number2022314964
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 12/21/2022
Supplement Date Received by Manufacturer12/21/2022
Initial Report FDA Received Date12/23/2022
Supplement Report FDA Received Date01/23/2023
Date Device Manufactured08/19/2020
Is the Device Labeled for Single Use? (Y/N) Yes
Usage of Device A
Patient Sequence Number1
Patient SexUnknown
-
-