ABBOTT QUADRA ALLURE CRT-P; IMPLANTABLE PACEMAKER PULSE GENERATOR Back to Search Results
Model Number PM3140
Medical Device Problem Code Pacemaker Found in Back-Up Mode (1440)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 12/12/2022
Type of Reportable Event Serious Injury
Event or Problem Description
During an in clinic follow-up, upon interrogation it was noted the device was in back up mode.The device was explanted and replaced to resolve the event.The patient was stable.
 
Additional Manufacturer Narrative
The reported event of device being found in backup mode was confirmed.The device was received with normal telemetry communication showing device in backup mode and output was not available.Backup mode occurred due to unknown power on reset while device was in the field.Longevity assessment found the device was operating at the elective replacement indicator (eri) consistent with normal usage.Additional findings: visual inspection of the header attachment area detected an anomaly between the epoxy header and titanium case.Feedthrough leak test was performed, indicating no sign of hermeticity breach.A manufacturing process anomaly may have occurred, which resulted in the header bonding anomaly.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
QUADRA ALLURE CRT-P
Common Device Name
IMPLANTABLE PACEMAKER PULSE GENERATOR
Manufacturer (Section D)
ABBOTT
15900 valley view court
sylmar CA 91342
Manufacturer (Section G)
ABBOTT
15900 valley view court
sylmar CA 91342
Manufacturer Contact
richard williamson
15900 valley view court
sylmar, CA 91342
MDR Report Key16043004
Report Number2017865-2022-50685
Device Sequence Number14761830
Product Code NIK
UDI-Device Identifier05414734507899
UDI-Public05414734507899
Combination Product (Y/N)N
Initial Reporter CountrySP
PMA/510(K) Number
P030035
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2015
Device Explanted Year2022
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 03/15/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date12/31/2016
Device Model NumberPM3140
Device Catalogue NumberPM3140
Device Lot NumberA000007986
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 12/12/2022
Supplement Date Received by Manufacturer02/22/2023
Initial Report FDA Received Date12/23/2022
Supplement Report FDA Received Date03/15/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured06/18/2015
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient SexUnknown
-
-