MAUDE Adverse Event Report: STRYKER NEUROVASCULAR CORK TARGET 360 SOFT 7MM X 30CM; DEVICE, NEUROVASCULAR EMBOLIZATION

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STRYKER NEUROVASCULAR CORK TARGET 360 SOFT 7MM X 30CM; DEVICE, NEUROVASCULAR EMBOLIZATION Back to Search Results
Model Number M0035477300
Medical Device Problem Codes Expulsion (2933); Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 12/08/2022
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that during aneurysm embolization procedure, the subject coil stretched while deploying inside the target lesion.Physician was unable to remove the subject coil from the patient anatomy.Additional medication was given to the patient due to this event.The procedure was completed successfully.No clinical consequences were reported to the patient due to this event.
 
Additional Manufacturer Narrative
The device is not available to the manufacturer.
 
Additional Manufacturer Narrative
Due to the automated manufacturing execution system (mes), there are controls in the manufacturing process to ensure the product met specifications upon release.The subject device is not available; therefore, functional testing as well as visual testing cannot be performed.The reported event is covered in the device directions for use (dfu).As well, the risk of the reported event is documented in the risk documentation and there are current controls to mitigate the risk of the as reported event.The reported complaint could not be confirmed, and it could not be definitively determined if the device failed to meet specifications because the product was not returned.Additional information received by the customer that the device prepared according to the instructions for use and continuous flushing was maintained throughout the procedure.While there are a number of potential causes for the reported issue, because review and analysis of available information failed to identify a definitive cause and the device was not returned, an assignable cause of undeterminable was assigned.
 
Event or Problem Description
It was reported that during aneurysm embolization procedure, the subject coil stretched while deploying inside the target lesion.Physician was unable to remove the subject coil from the patient anatomy.Additional medication was given to the patient due to this event.The procedure was completed successfully.No clinical consequences were reported to the patient due to this event.Additional information received on 17-jan-2023, clarified that the subject coil was also protruding from the aneurysm to the artery.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TARGET 360 SOFT 7MM X 30CM
Common Device Name
DEVICE, NEUROVASCULAR EMBOLIZATION
Manufacturer (Section D)
STRYKER NEUROVASCULAR CORK
ida industrial estate
model farm road
cork NA
EI  NA
Manufacturer (Section G)
STRYKER NEUROVASCULAR CORK
ida industrial estate
model farm road
cork NA
EI   NA
Manufacturer Contact
tara lopez
47900 bayside parkway
fremont, CA 94538
5104132500
MDR Report Key16043078
Report Number3008881809-2022-00652
Device Sequence Number4897082
Product Code HCG
UDI-Device Identifier04546540676580
UDI-Public04546540676580
Combination Product (Y/N)N
Initial Reporter CountryFR
PMA/510(K) Number
K153658
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 03/03/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date12/06/2024
Device Model NumberM0035477300
Device Catalogue NumberM0035477300
Device Lot Number23352810
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 12/09/2022
Supplement Date Received by Manufacturer02/10/2023
Initial Report FDA Received Date12/23/2022
Supplement Report FDA Received Date03/03/2023
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured12/08/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient SexUnknown
-
-