SMITH & NEPHEW, INC. META-NAIL TIBIAL 10MM X 37CM; NAIL, FIXATION, BONE Back to Search Results
Model Number 71655137
Medical Device Problem Code Delivered as Unsterile Product (1421)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 01/27/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The device was not returned for evaluation but the pictures were reviewed, and revealed the presence of an unintended particle in the nail.A review of the production order did not reveal a manufacturing abnormality that could have caused or contributed to the reported incident.A review of complaint history for the part number over the past 12 months and for the batch number based on historical data of the device did not reveal similar events for the listed device.A review of the risk management file revealed this failure mode was previously identified.The anticipated risk level is still adequate.A historical review concluded that there are no prior actions related to this product and event.At this time, we have no reason to suspect that the product failed to meet any product specifications at the time of manufacture.Factor that could contribute to the reported event include but not limited to a grindstone jammed in the device during the surface finishing process.The contribution of the device to the reported event could be corroborated since the photos confirm the presence of an unintended particle in the nail.Based on this investigation, the need for corrective action is not indicated.Should the device or additional information be received, the complaint will be reopened.No further investigation is warranted for this complaint; however, we will continue to monitor for future complaints and investigate as necessary.We consider this investigation closed.Internal complaint reference number: (b)(4).
 
Event or Problem Description
It was reported that, during an internal fixation surgery, a rock was found within the meta-nail tibial 10mm x 37cm when opened.It is unknown how surgery was finished and if it was delayed; however, patient harm was not indicated.
 
Additional Manufacturer Narrative
H3, h6: the device was not returned for evaluation but the pictures were reviewed and revealed the presence of an unintended particle in the nail.A review of the production order did not reveal a manufacturing abnormality that could have caused or contributed to the reported incident.A review of complaint history for the part number over the past 12 months and for the batch number based on historical data of the device did not reveal similar events for the listed device.A review of the risk management file revealed this failure mode was previously identified.The anticipated risk level is still adequate.A historical review concluded that there are no prior actions related to this product and event.At this time, we do have evidence to suspect that the product failed to meet any product specifications at the time of manufacture.An investigation made by the quality engineering team revealed that worn grinding stones changed too late could be a factor for the unintended particle in the nail.With the results of this investigation the root cause of this event could not be determined.However, certain process specifications may not have been followed in the manufacturing process.Based on this investigation, the need for corrective actions is indicated.Inspection plans and standard rework process are going to be improve in the manufacturing process.Should the device or additional information be received, the complaint will be reopened.No further investigation is warranted for this complaint; however, we will continue to monitor for future complaints and investigate as necessary.We consider this investigation closed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
META-NAIL TIBIAL 10MM X 37CM
Common Device Name
NAIL, FIXATION, BONE
Manufacturer (Section D)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
Manufacturer (Section G)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
Manufacturer Contact
holly topping
7000 west william cannon drive
austin, TX 78735
5123913905
MDR Report Key16043354
Report Number1020279-2022-05030
Device Sequence Number20160312
Product Code JDS
UDI-Device Identifier03596010585592
UDI-Public03596010585592
Combination Product (Y/N)N
Initial Reporter StateGA
Initial Reporter CountryUS
PMA/510(K) Number
K061019
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source User Facility,Company Representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 02/20/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number71655137
Device Catalogue Number71655137
Device Lot Number21FT51953
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 01/27/2022
Supplement Date Received by Manufacturer02/17/2023
Initial Report FDA Received Date12/23/2022
Supplement Report FDA Received Date02/20/2023
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured06/24/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-