• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BECTON, DICKINSON AND COMPANY, BD BIOSCIENCES BD FACS¿ SAMPLE PREP ASSISTANT III; SEE H.10

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BECTON, DICKINSON AND COMPANY, BD BIOSCIENCES BD FACS¿ SAMPLE PREP ASSISTANT III; SEE H.10 Back to Search Results
Model Number 647205
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 12/08/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
A device evaluation and/or device history review is anticipated but is not complete.Upon completion, a supplemental report will be filed.Common device name: automated pipetting, diluting and specimen processing workstations for flow cytometric analysis.Initial reporter phone: (b)(6).
 
Event or Problem Description
It was reported that bd facs¿ sample prep assistant iii leakage contained within the instrument.The following information was provided by the initial reporter leakage contained within the instrument in a customer accessible location.Leaked fluid: sheath fluid.The source was non-waste line.Customer was not exposed to blood or bodily fluids.The customer suffered no physical harm as result of the leak.
 
Event or Problem Description
It was reported that bd facs¿ sample prep assistant iii leakage contained within the instrument.The following information was provided by the initial reporter leakage contained within the instrument in a customer accessible location.Leaked fluid: sheath fluid.The source was non-waste line.Customer was not exposed to blood or bodily fluids.The customer suffered no physical harm as result of the leak.
 
Additional Manufacturer Narrative
After further evaluation of the complaint, it has been determined that the previously submitted report 2916837-2022-00370 was sent in error.There was no report of serious injury, medical intervention, or reportable device malfunction leakage of non-pressurized fluids are unlikely to lead to a serious injury or death due to the passive nature is not likely to lead contact of the user or other person.Therefore this is not considered to be a reportable malfunction bdb has monitored its mdr reporting, and after review, has determined through valid data that: the relevant malfunctions have not caused or contributed to further deaths or serious injuries for two years, and the likelihood of another death or serious injury as a result of the relevant malfunctions is remote.Accordingly, bdb has amended its mdr reporting guidelines to discontinue further presumptive reporting of malfunctions associated with liquid leakage and splash complaints.Mfr.#2916837-2022-00370 was filed under previous guidance and is now corrected to be a non-reportable event.        the aforementioned amendment to the bdb mdr reporting guidelines was based on the following rationale: to evaluate the impact of this proposed change, an assessment of the complaint and adverse event reporting history was performed for malfunctions associated with serious injuries or deaths related to liquid leakage and splash events.The assessment confirms there have been no reports of serious injury or death related to liquid leakage or splash events for the last two years ((b)(6) 2020 thru (b)(6)2022).    we also note that the likelihood of another death or serious injury as a result of the relevant malfunctions is remote.Specifically, a liquid leakage or splash of a chemical or potentially biohazardous fluid would be readily apparent to the user and is unlikely to come in contact with eyes, mouth, other mucous membrane, and non-intact skin, or cause parenteral contact with blood or other potentially infectious materials.Users are also instructed to follow universal precautions including wearing ppe per the products user documentation, and the requirements in the osha bloodborne pathogen guideline.As such, the risk of serious injury or death occurring related to malfunctions causing liquid leakage or splash events would be improbable.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BD FACS¿ SAMPLE PREP ASSISTANT III
Common Device Name
SEE H.10
Manufacturer (Section D)
BECTON, DICKINSON AND COMPANY, BD BIOSCIENCES
2350 qume drive
san jose CA 95131
Manufacturer (Section G)
BECTON, DICKINSON AND COMPANY, BD BIOSCIENCES
2350 qume drive
san jose CA 95131
Manufacturer Contact
phillip emmert
9450 south state street
sandy, UT 84070
8015296192
MDR Report Key16043434
Report Number2916837-2022-00370
Device Sequence Number6238176
Product Code PER
UDI-Device Identifier00382906472052
UDI-Public00382906472052
Combination Product (Y/N)N
Initial Reporter CountryIT
PMA/510(K) Number
K131301
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Other,User Facility
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 03/28/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Other
Device Model Number647205
Device Catalogue Number647205
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 12/15/2022
Supplement Date Received by Manufacturer03/28/2023
Initial Report FDA Received Date12/23/2022
Supplement Report FDA Received Date04/05/2023
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured05/19/2017
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-