MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION ACCOLADE DR; PACEMAKER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION ACCOLADE DR; PACEMAKER Back to Search Results
Model Number L301
Device Problem Defective Device (2588)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 11/22/2022
Event Type  malfunction  
Manufacturer Narrative
Upon receipt at our post market quality assurance laboratory, this device was thoroughly inspected and analyzed.A high current drain was detected, but the battery still supported full device function.Detailed analysis confirmed the identified high current drain was a result of compromised capacitors.This resulted in the observed premature battery depletion.It was determined that the capacitors were compromised due to the presence of excess hydrogen in the device case.Boston scientific issued a field safety notice regarding a subset of devices in the accolade pacemaker family that have a potential of exhibiting this behavior.This particular device is included in the hydrogen induced premature depletion advisory population.
 
Event Description
It was reported that this pacemaker was explanted without allegations from the field.It was subsequently returned for analysis.No adverse patient effects were reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ACCOLADE DR
Type of Device
PACEMAKER
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
4100 hamline avenue north
saint paul MN 55112
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
4100 hamline avenue north
saint paul MN 55112 5798
Manufacturer Contact
timothy degroot
4100 hamline avenue north
saint paul, MN 55112
6515826168
MDR Report Key16043494
MDR Text Key308004922
Report Number2124215-2022-54812
Device Sequence Number1
Product Code LWP
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
N970003/S167
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Company Representative
Reporter Occupation Other
Remedial Action Notification
Type of Report Initial
Report Date 12/23/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Expiration Date05/11/2017
Device Model NumberL301
Device Catalogue NumberL301
Device Lot Number708638
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/22/2022
Initial Date Manufacturer Received 12/20/2022
Initial Date FDA Received12/23/2022
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured05/11/2015
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age68 YR
Patient SexMale
-
-