Model Number 6660 |
Device Problems
Battery Problem (2885); No Apparent Adverse Event (3189)
|
Patient Problems
Inadequate Pain Relief (2388); No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 11/29/2022 |
Event Type
Injury
|
Manufacturer Narrative
|
Date of event is estimated.
|
|
Event Description
|
Related manufacturer report number: 1627487-2022-07059.It was reported that the patient's ipgs were inoperable.It is unknown when the patient lost therapy.Surgical intervention was undertaken wherein the ipgs were explanted and replaced.Effective therapy was established post operatively.
|
|
Manufacturer Narrative
|
The results of the investigation are inconclusive since the device was not returned for analysis.Based on the information received, the cause of the reported incident could not be conclusively determined.
|
|
Event Description
|
Additional information was received that the ipg met or exceeded 75% expected longevity.There is no device malfunction.
|
|