| Model Number |
CD2257-40Q |
| Medical Device Problem Code |
Premature Discharge of Battery (1057)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
12/07/2022
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
It was reported that a patient presented remotely on (b)(6) 2022 with their implantable cardioverter defibrillator (icd) at elective replacement indicator (eri), despite having had 9 months of remaining longevity during the last check in (b)(6) 2022.Premature battery depletion was suspected.The patient's icd was explanted and replaced on (b)(6) 2022.The patient was stable throughout.
|
| |
|
Additional Manufacturer Narrative
|
|
Premature battery depletion was confirmed by analysis.No sources of high current were noted.The cause of the premature battery depletion was consistent with lithium (li) cluster formation.From these analyses, in the absence of high current draw, it is probable that the premature battery depletion was caused by a lithium cluster induced short circuit.Li clusters are a known depletion mechanism for these advisory products that has been investigated and associated with a field action in october 2016.
|
| |
|