| Model Number |
8220 ALARIS SPO2 MODULE |
| Medical Device Problem Code |
No Apparent Adverse Event (3189)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Note: this event was inadvertently reported under incorrect product registration number (manufacturer report number 9616066-2021-00007).This mdr resubmission now reflects the correct product registration number.The customer reported problem was confirmed.A review of the device history record in sap for sn: (b)(4) was performed from the date of the manufacture to date of the release of product, which confirmed that this device was not involved in a production failure, and product was returned for servicing which correlates to the customer reported issue.A review of the complaint history record in the trackwise was performed for the sn: (b)(4) which confirmed no similar complaints with the same or related failure mode.The customer stated that there was no patient involvement.
|
| |
|
Event or Problem Description
|
|
Error code 471.6211.Communication failure in spo2 board.On (b)(6) 2020 07:43:43 (b)(6).Shipping & billing addresses confirmed.There was no patient involvement.
|
| |
|
Additional Manufacturer Narrative
|
|
Addition: updated adverse event codes (h6).
|
| |
|
Event or Problem Description
|
|
(b)(4).
|
| |
|
Search Alerts/Recalls
|