VENUM20100 |
| Medical Device Problem Codes |
Misfire (2532); Activation Failure (3270)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
11/29/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
A voluntary recall has been initiated for the venovo¿ venous stent system 9f which was product catalog/lot number specific.Reportedly the proximal end of the stent does not immediately expand upon deployment.The proximal end of the stent remains connected to the delivery system.As a result of the field action, this event is being reported as a malfunction reportable event.As the lot number for the device was provided, a review of the device history records is currently being performed.The return of the sample is pending.The investigation of the reported event is currently underway.Device pending return.
|
| |
|
Event or Problem Description
|
|
It was reported that during a stent placement procedure, the distal portion of the stent allegedly would not come out of the deployment catheter.There was no reported patient injury.
|
| |
|
Event or Problem Description
|
|
It was reported that during a stent placement procedure, the distal portion of the stent allegedly would not come out of the deployment catheter.There was no reported patient injury.
|
| |
|
Additional Manufacturer Narrative
|
|
H10: manufacturing review: the lot history records of this lot were reviewed with special attention to the manufacturing and inspection of this product and the product was found to have met the specification prior to shipment.Based on the information available it is not reasonably suggested that a manufacturing process may have caused or contributed to the reported issue.Investigation summary: the sample was not returned for evaluation and image was not provided for review.Therefore, the investigation is inconclusive for the reported deployment issue.The definitive root cause could not be determined based upon available information.Labeling review: in reviewing the relevant labeling for this product the potential issue was found addressed.The instructions for use state: 'if excessive force is felt during stent deployment, do not force the delivery system.Remove the delivery system and replace with a new unit.', and 'examine the stent system for any damage.If it is suspected that the sterility or performance of the stent system has been compromised, the device must not be used.' the instructions for use further state: 'predilation of chronic lesions with a balloon dilatation catheter is recommended.', and under materials required the instructions for use state: '10f introducer for stent diameters of 16 mm, 18 mm and 20 mm 0.035 inch diameter guidewire'.Holding and handling of the system throughout deployment was found sufficiently described.H11: section a through f ¿ the information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant/reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
|
| |
|