• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITH & NEPHEW, INC. GRSPR CUPD 3.4MM; ACCESSORIES,ARTHROSCOPIC

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITH & NEPHEW, INC. GRSPR CUPD 3.4MM; ACCESSORIES,ARTHROSCOPIC Back to Search Results
Model Number 011014
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 11/25/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Internal complaint reference: (b)(4).
 
Event or Problem Description
It was reported that during a shoulder arthroscopic rotator cuff repair, the grasper was broken, the jaws failed to close and lock.It was removed from the patient and inspected.The scrub tech tried to open and close the grasper and the top of the jaws and the articulation rod came apart.The procedure was completed with a s+n back-up device.There was no surgical delay and no further complications were reported.
 
Additional Manufacturer Narrative
H10: internal complaint reference (b)(4).H3, h6: the reported device was received for evaluation.A visual inspection of the returned device found that it is not in its original packaging.The grasper is fractured at the distal tip and the shaft is disconnected from the handle at the proximal end of the shaft.The set screw is disengaged.A functional evaluation could not be performed because the device is broken into several distinct pieces.A review of device records showed there were no indications to suggest that the product did not meet manufacturing specification upon release for distribution.A complaint history review found similar reported events.The instructions for use was reviewed and found to include conditions of off label use and technique specifics, as well as precautions and warnings related to the use of the device.A risk management review found that the reported failure and/or harm was documented appropriately, and there were no indications to suggest the anticipated risk is not adequate.The root cause was associated with unintended use of the device.Factors that could have contributed to the reported event include excessive force on the device, attempted correction of a damaged device, or an inadvertent impact event inconsistent with normal use.No containment or corrective actions are recommended at this time.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
GRSPR CUPD 3.4MM
Common Device Name
ACCESSORIES,ARTHROSCOPIC
Manufacturer (Section D)
SMITH & NEPHEW, INC.
130 forbes blvd.
mansfield MA 02048
Manufacturer (Section G)
SMITH & NEPHEW, INC.
130 forbes blvd.
mansfield MA 02048
Manufacturer Contact
holly topping
7000 west william cannon drive
austin, TX 78735
5123913905
MDR Report Key16048568
Report Number1219602-2022-02043
Device Sequence Number12319655
Product Code NBH
UDI-Device Identifier03596010022936
UDI-Public03596010022936
Combination Product (Y/N)N
Initial Reporter StateCA
Initial Reporter CountryUS
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 01/02/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number011014
Device Catalogue Number011014
Device Lot Number50967888
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/19/2022
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 12/05/2022
Supplement Date Received by Manufacturer12/22/2023
Initial Report FDA Received Date12/26/2022
Supplement Report FDA Received Date01/02/2024
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured10/06/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Patient SexUnknown
-
-