| Model Number |
71174904 |
| Medical Device Problem Code |
Fracture (1260)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
11/30/2022
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Internal complaint reference: (b)(4).
|
| |
|
Event or Problem Description
|
|
It was reported that, during a fibula and tibia osteosynthesis, upon fastening the plate, a 1.5mm provisional fixation pin 20mm broke off while it was removed and a part of it remained inside the patient.This part was successfully retrieved from the patient's incision.Surgery was resumed, after a non-significant delay, with a smith and nephew back-up device.The patient was reported in an stable condition upon procedure completion.
|
| |
|
Additional Manufacturer Narrative
|
|
H6: health effect - impact code.Section h3, h6: the device was not returned for evaluation; therefore, a device analysis could not be performed.The clinical/medical investigation concluded that, as of the date of this medical investigation, the supporting clinical documentation has not been provided; therefore, no clinical factors could be concluded to have contributed to the reported event.Patient impact beyond the reported broken device, modified surgical procedure for retrieval with a 0-30 minute surgical delay could not be determined.No further medical assessment can be rendered at this time.Device batch number was not provided, thus, an evaluation of the manufacturing records could not be performed.A review of complaint history for the previous 12 months did not reveal similar events for the listed device.A review of the instructions for use file is not applicable for this instrument.A review of the risk management file revealed this failure mode was previously identified.The anticipated risk level is still adequate.A historical review concluded that there are no prior actions related to this product and event.At this time, we have no reason to suspect that the product failed to meet any product specifications at the time of manufacture.With the results of this investigation the root cause of this event could not be determined.This is a single use device, surgical technique or damage from misuse are likely potential factors that could contribute to the reported event.Based on this investigation, the need for corrective action is not indicated.Should the device or additional information be received, the complaint will be reopened.No further investigation is warranted for this complaint; however, we will continue to monitor for future complaints and investigate as necessary.We consider this investigation closed.
|
| |
|
Additional Manufacturer Narrative
|
|
H3, h6.The associated device was returned and evaluated.The visual inspection revealed the pin fractured into two pieces.Both pieces were returned.The clinical/medical investigation concluded that, per complaint details, the device ¿broke off¿ during use; however, the piece was ¿successfully retrieved¿ from incision.Reportedly, surgery was resumed, after a non-significant delay, with a smith and nephew back-up device with patient in stable condition upon procedure completion/¿patient¿s health was not affected¿.As of the date of this medical investigation, the supporting clinical documentation has not been provided; therefore, no clinical factors could be concluded to have contributed to the reported event.Patient impact beyond the reported broken device, modified surgical procedure for retrieval with a 0-30 minute surgical delay could not be determined.No further medical assessment can be rendered at this time.A review of the production order did not reveal a manufacturing abnormality that could have caused or contributed to the reported incident.A review of complaint history for the part number over the past 12 months and for the batch number based on historical data of the device did not reveal similar events for the listed device.A review of instructions for use for care, maintenance, cleaning and sterilization of smith & nephew orthopedics devices revealed that visually inspecting for damage prior to use is indicated.A review of the risk management file revealed this failure mode was previously identified.The anticipated risk level is still adequate.A historical review concluded that there are no prior actions related to this product and event.At this time, we have no reason to suspect that the product failed to meet any specifications at the time of manufacture.This is a single use device, surgical technique or damage from misuse are likely potential factors that could contribute to the reported event.Based on this investigation, the need for corrective action is not indicated.Should additional information be received, the complaint will be reopened.No further investigation is warranted for this complaint; however, we will continue to monitor for future complaints and investigate as necessary.We consider this investigation closed.
|
| |
|