| Model Number |
72201882 |
| Medical Device Problem Code |
Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Codes |
Hyperplasia (1906); Swelling/ Edema (4577)
|
| Date of Event |
12/02/2022
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
It was reported that six months after a left pinky repair surgery, where a spyromite was used, the patient experienced swelling on the finger, on december 2nd a revision surgery was performed, where it was found exudation around the tendon suture knot, pterygium-like necrosis around the staple fixation suture with thread and inflammatory hyperplasia of the surrounding tissue; and therefore the sutures were removed.Patient´s currently fine.As this was noticed after the original procedure, this was finished with the reported device.No further complications were reported, as the reported ones, were solved after the revision surgery.
|
| |
|
Additional Manufacturer Narrative
|
|
Internal complaint reference: (b)(4).
|
| |
|
Additional Manufacturer Narrative
|
|
H10 h6: a review of device records showed there were no indications to suggest that the product did not meet manufacturing specifications upon release for distribution.A complaint history review found no similar reported events.The instructions for use was reviewed and found to include conditions of off label use and technique specifics, as well as precautions and warnings related to the use of the device.A risk management review found that the reported failure and/or harm was documented appropriately, and there were no indications to suggest the anticipated risk is not adequate.A clinical review states the provided intraoperative photo was reviewed and supports the reported event, a surgically incised wound with a piece of suture noted in the wound bed.The wound bed also appears to be pale to grey in color, however the wound borders cannot be properly assessed from the provided image.Additionally, the photo does not provide insight into the reported event cause.No wound or serum blood cultures were provided for review.All documents and images provided as of this date have been reviewed and considered and unless noted do not contribute to the clinical/medical investigation.Based on the information provided, the root cause of the reported left little finger swelling, and subsequent revision cannot be definitively concluded but infection is a known postoperative complication.Although the operative findings for the revision was provided, the requested operative report surgical technique for the initial surgery was not provided for review in this medical investigation.Therefore, no further clinical/medical assessment is warranted.The reported adverse event was confirmed, however a device deficiency was not identified since the product was not returned.Please refer to the instructions for use for recommendations on proper use of the device and potential troubleshooting methods to prevent future reoccurrence of the reported event.No containment or corrective actions are recommended at this time.
|
| |
|
Search Alerts/Recalls
|
|
|