MAUDE Adverse Event Report: STRYKER TRAUMA KIEL FEMORAL NAIL GT, RIGHT T2 ALPHA FEMUR ANTEGRADE Ø10X360MM; ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STRYKER TRAUMA KIEL FEMORAL NAIL GT, RIGHT T2 ALPHA FEMUR ANTEGRADE Ø10X360MM; ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES Back to Search Results
Model Number 2333-1036S
Medical Device Problem Codes Break (1069); Fracture (1260)
Health Effect - Clinical Codes Failure of Implant (1924); Nonunion/Delayed-union Bone Fracture (2369)
Date of Event 12/02/2022
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Once the investigation has been completed any additional information will be reported in a supplemental report.
 
Event or Problem Description
It was reported that a nail broke and a revision surgery was performed.Additionally a nonunion was reported.
 
Additional Manufacturer Narrative
The reported event could be confirmed, since the device was returned for evaluation and matches the alleged failure mode.The device inspection revealed the following: the received nail was found to be broken from almost mid portion i.E.Shaft of the nail.Overall, the fracture pattern reveals a fatigue failure evident by the appearance of lines of rest.Just opposite to the origin point of the breakage, plastic deformation of edges, evident by shiny surface, could be observed indicating towards excessive loading over a considerable period of time.A review of the device history for the reported lot did not indicate any abnormalities.No corrective actions are required at this time.No indications of material, manufacturing or design related problems were found during the investigation.A review of the labeling did not indicate any abnormalities.Based on the evidences from the device inspection and the available information like a potential non-union (cannot be confirmed without x-ray examination), the most likely cause of the failure is patient related.The fact that the nail broke almost after 8 months indicates towards a potential non-union.It cannot be excluded that a suboptimal fracture reduction/treatment could also be a reason for non-union and subsequent nail breakage, but x-rays are needed to confirm that.If any additional information is provided, the investigation will be reassessed.
 
Event or Problem Description
It was reported that a nail broke and a revision surgery was performed.Additionally a nonunion was reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FEMORAL NAIL GT, RIGHT T2 ALPHA FEMUR ANTEGRADE Ø10X360MM
Common Device Name
ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES
Manufacturer (Section D)
STRYKER TRAUMA KIEL
prof. kuentscher-strasse 1-5
schoenkirchen/kiel D-242 32
GM  D-24232
Manufacturer (Section G)
STRYKER TRAUMA KIEL
prof. kuentscher-strasse 1-5
schoenkirchen/kiel D-242 32
GM   D-24232
Manufacturer Contact
anna jusinski
325 corporate drive
mahwah, NJ 07430
2018315000
MDR Report Key16048862
Report Number0009610622-2022-00601
Device Sequence Number12326789
Product Code HSB
UDI-Device Identifier07613327285994
UDI-Public07613327285994
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K203819
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2022
Device Explanted Year2022
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 03/01/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Model Number2333-1036S
Device Catalogue Number23331036S
Device Lot NumberK08E770RA
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 12/02/2022
Supplement Date Received by Manufacturer02/02/2023
Initial Report FDA Received Date12/26/2022
Supplement Report FDA Received Date03/01/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured01/28/2020
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age50 YR
Patient SexFemale
-
-