• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: HOYA CORPORATION PENTAX TOKYO OFFICE PENTAX; VIDEO NASOLARYNGOSCOPE (FOR VIVIDEO)

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

HOYA CORPORATION PENTAX TOKYO OFFICE PENTAX; VIDEO NASOLARYNGOSCOPE (FOR VIVIDEO) Back to Search Results
Model Number VNL9-CP
Device Problem Break (1069)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 12/14/2022
Event Type  malfunction  
Manufacturer Narrative
We checked the returned unit and confirmed that the ccd driver pcb as defective.Based on the result, we concluded that it was caused due to the excessive force applied on the ccd driver pcb.In addition, we confirmed that the insertion flexible tube (ift) coating damage, the distal body leak, and the light guide cable cut; however, they are not the main cause and/or irrelevant to the alleged complaint.Based on the technical report, "hr-rpt-0586(image failure)" and/or the risk analysis results, it was evaluated to submit mdr.
 
Event Description
The time of event is unknown.There was no report of patient harm.Video image failure.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PENTAX
Type of Device
VIDEO NASOLARYNGOSCOPE (FOR VIVIDEO)
Manufacturer (Section D)
HOYA CORPORATION PENTAX TOKYO OFFICE
tsutsujigaoka 1-1-110
akishima-shi, tokyo 196-0 012
JA  196-0012
Manufacturer (Section G)
HOYA CORPORATION PENTAX TOKYO OFFICE
tsutsujigaoka 1-1-110
akishima-shi, tokyo 196-0 012
JA   196-0012
Manufacturer Contact
william goeller
3 paragon drive
montvale, NJ 07645
MDR Report Key16049197
MDR Text Key308460455
Report Number9610877-2022-61571
Device Sequence Number1
Product Code EOB
Combination Product (y/n)N
Reporter Country CodeIT
PMA/PMN Number
K171011
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional
Reporter Occupation Physician
Type of Report Initial
Report Date 12/27/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/27/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberVNL9-CP
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/14/2022
Is the Reporter a Health Professional? Yes
Date Manufacturer Received12/14/2022
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-