• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: OSTEOMED, LLC 1.6MM X 5MM AUTO-DRIVE SCREW; SCREW, FIXATION, BONE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

OSTEOMED, LLC 1.6MM X 5MM AUTO-DRIVE SCREW; SCREW, FIXATION, BONE Back to Search Results
Model Number 211-1605
Medical Device Problem Code Patient Device Interaction Problem (4001)
Health Effect - Clinical Code Discomfort (2330)
Date of Event 10/04/2022
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
The investigation is currently pending, and a follow up report will be submitted upon completion of the investigation.Related reports: 2027754-2022-00142, 2027754-2022-00143, 2027754-2022-00144, 2027754-2022-00145, 2027754-2022-00146, 2027754-2022-00147, 2027754-2022-00148, 2027754-2022-00149, 2027754-2022-00150, 2027754-2022-00152, 2027754-2022-00153, 2027754-2022-00154, 2027754-2022-00155, 2027754-2022-00156, 2027754-2022-00157, 2027754-2022-00158, 2027754-2022-00159, 2027754-2022-00160, 2027754-2022-00161, 2027754-2022-00162, 2027754-2022-00163, 2027754-2022-00164, 2027754-2022-00165, 2027754-2022-00166, 2027754-2022-00167, 2027754-2022-00168, 2027754-2022-00169, 2027754-2022-00170, 2027754-2022-00171, 2027754-2022-00172, 2027754-2022-00173, 2027754-2022-00174, 2027754-2022-00175, 2027754-2022-00176, 2027754-2022-00177, 2027754-2022-00178, 2027754-2022-00179, 2027754-2022-00180, 2027754-2022-00181, 2027754-2022-00182, 2027754-2022-00183.2027754-2022-00184, 2027754-2022-00185, 2027754-2022-00186, 2027754-2022-00187, 2027754-2022-00188, 2027754-2022-00189, 2027754-2022-00190, 2027754-2022-00191, 2027754-2022-00192, 2027754-2022-00193, 2027754-2022-00194, 2027754-2022-00195, 2027754-2022-00196, 2027754-2022-00197, 2027754-2022-00198.
 
Event or Problem Description
The patient was implanted on (b)(6) 2022 with no issues post-op or during clinical follow up until the patient started reporting discomfort around (b)(6) 2022 (event date) at which time imaging suggested the screw on the left side plate appeared out of the usual location.Exploratory surgery was performed on (b)(6) 2022 and found loosening of the plate and "slight movement of the distal stump".The patient was then referred to the (b)(6) for emergency removal and refixation which took place on (b)(6) 2022.The impression was that the mandibullary bone had reabsorbed around the screws and granulation tissue had formed at the site.Bacterial culture of material collected revealed it was negative.It was reported "they will be forwarding the screws removed from the patient's second surgery"- presumably, the devices explanted during the revision on (b)(6) 2022.It was also reported the patient might be diagnosed with metallosis (clinical hypothesis), and it could be related to plates and screws implanted.No further information has been provided.Report 10 of 57 for this event.
 
Additional Manufacturer Narrative
The device history records (dhrs) for the provided lot numbers were individually reviewed.All inspections of all lots passed, and no issues were identified in the dhr review.A two-year review of complaints were performed, and zero (0) complaints for this issue (implant loosening) were noted.The literature contains the proper warnings.Devices/products were not returned for further evaluation; therefore, the root cause of the concern could not be determined.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
1.6MM X 5MM AUTO-DRIVE SCREW
Common Device Name
SCREW, FIXATION, BONE
Manufacturer (Section D)
OSTEOMED, LLC
3885 arapaho rd
addison TX 75001
Manufacturer (Section G)
OSTEOMED, LLC
3885 arapaho rd
addison TX 75001
Manufacturer Contact
ellie wood
3885 arapaho rd
addison, TX 75001
9726774600
MDR Report Key16049547
Report Number2027754-2022-00151
Device Sequence Number17753046
Product Code HWC
UDI-Device Identifier00845694002370
UDI-Public(01)00845694002370(10)1165972(30)1(11)220113
Combination Product (Y/N)N
Initial Reporter CountryBR
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2022
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Distributor
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date (Section B) 01/12/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Model Number211-1605
Device Catalogue Number211-1605
Device Lot Number1165972
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 12/15/2022
Supplement Date Received by Manufacturer01/12/2023
Initial Report FDA Received Date12/27/2022
Supplement Report FDA Received Date01/13/2023
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient SexFemale
-
-