• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTRONIC PUERTO RICO OPERATIONS CO. 640G INSULIN PUMP MMT-1712K; PUMP, INFUSION, INSULIN, TO BE USED WITH INVASIVE GLUCOSE SENSOR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC PUERTO RICO OPERATIONS CO. 640G INSULIN PUMP MMT-1712K; PUMP, INFUSION, INSULIN, TO BE USED WITH INVASIVE GLUCOSE SENSOR Back to Search Results
Model Number MMT-1712K
Device Problem Display or Visual Feedback Problem (1184)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 12/08/2022
Event Type  malfunction  
Event Description
Information received by medtronic indicated that the customer called to report partial display.Blood glucose value at the time of event was 117 mg/dl.Troubleshooting was performed.No harm requiring medical intervention was reported.The insulin pump will be returned for analysis.The customer will discontinue use of the device.
 
Manufacturer Narrative
(b)(4).The insulin pump involved in this event is the ngp 640g insulin pump which is not marketed in the united states.However, the device is similar to the ngp insulin pump, which is marketed in the united states currently it is unknown whether or not the device may have caused or contributed to the event as no product has been returned.The device will be returned for analysis and further information will follow once the analysis has been completed.No conclusion can be drawn at this time.Medtronic, inc.(medtronic) is submitting this report to comply with 21 c.F.R.Part 803, the medical device reporting regulation.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information in the time allotted and has provided as much information as is available to the company as of the submission date this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employees caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any "defects" or has "malfunctioned".These words are included in the fda 3500a form and are fixed items for selection created by the fda, to categorize the type of event solely for the purpose of reporting pursuant to part 803.Medtronic objects to the use of these words and others like it because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.
 
Event Description
Correction summary: the device will be returned for analysis.
 
Manufacturer Narrative
On (b)(6) 2022 customer called in with the following concern: customer calls reporting partial display.P-cap locked in place properly during testing.Unit passed displacement test however, unable to complete a self test due to multiple lines across screen.In this condition, multiple traces on the lcd glass between the lcd controller and the lcd image had been broken, preventing certain horizontal or vertical lines of information from displaying.Unit also received with cracked case (battery tube) and cracked battery tube threads.In conclusion customer allegations were confirm during visual inspection when multiple lines across screen were noted due to broken traces.Medtronic, inc.(medtronic) is submitting this report to comply with 21 c.F.R.Part 803, the medical device reporting regulation.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information in the time allotted and has provided as much information as is available to the company as of the submission date this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employees caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any "defects" or has "malfunctioned".These words are included in the fda 3500a form and are fixed items for selection created by the fda, to categorize the type of event solely for the purpose of reporting pursuant to part 803.Medtronic objects to the use of these words and others like it because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
640G INSULIN PUMP MMT-1712K
Type of Device
PUMP, INFUSION, INSULIN, TO BE USED WITH INVASIVE GLUCOSE SENSOR
Manufacturer (Section D)
MEDTRONIC PUERTO RICO OPERATIONS CO.
ceiba norte ind. park #50 road
juncos 00777 -386
Manufacturer (Section G)
MEDTRONIC PUERTO RICO OPERATIONS CO.
ceiba norte ind. park #50 road
juncos 00777 -386
Manufacturer Contact
tricha miles
ceiba norte ind. park #50 road
juncos 00777--386
7635140379
MDR Report Key16049922
MDR Text Key306704760
Report Number2032227-2022-388381
Device Sequence Number1
Product Code OYC
Combination Product (y/n)N
Reporter Country CodeBR
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Company Representative
Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date 03/27/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/27/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Model NumberMMT-1712K
Device Catalogue NumberMMT-1712K
Device Lot NumberHG509RB
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Manufacturer Received03/03/2023
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured01/12/2021
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-