• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: WILLIAM COOK EUROPE FROVA INTUBATING INTRODUCER; BTR TUBE, TRACHEAL (W/WO CONNECTOR)

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

WILLIAM COOK EUROPE FROVA INTUBATING INTRODUCER; BTR TUBE, TRACHEAL (W/WO CONNECTOR) Back to Search Results
Model Number G48303
Medical Device Problem Codes Component Incompatible (1108); Flaked (1246); Off-Label Use (1494)
Health Effect - Clinical Code Foreign Body In Patient (2687)
Date of Event 12/08/2022
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Manufacturer ref# (b)(4).Blank fields on this form indicate the information is unknown or unavailable.G4) similar to device marketed under pma/510(k): k161813.Investigation is still in progress.This report is required by the fda under 21 cfr part 803.This report is based on unconfirmed information submitted by others.Neither the submission of this report nor any statement made in it is intended to be an admission that any cook device is defective or malfunctioned; that a death or serious injury occurred; or that any cook device caused or contributed to; or is likely to cause or contribute to a death or serious injury if a malfunction occurred.
 
Event or Problem Description
Description of event according to initial reporter: use of the frova chuck.Fibroscopic control to see the correct position of the right 41fr s selective airway, blue chuck fragment fount in patients bronchus.Patient outcome: current state of the patient: extraction of the fragment using the fiberscope.Increased risk of infection.
 
Event or Problem Description
No additional information regarding the patient and/or event has been received since the previous medwatch report was sent.
 
Additional Manufacturer Narrative
Manufacturers ref# (b)(4).This is a correction due to lot#, which was mistakenly included in the previous report.Cook medical will continue to monitor for similar events.This report is required by the fda under 21cfr part 803.This report is based on unconfirmed information submitted by others.Neither the submission of this report nor any statement made in it is intended to be an admission that any cook device is defective or malfunctioned; that a death or serious injury occurred; or that any cook device caused or contributed to; or is likely to cause or contribute to a death or serious injury if a malfunction occurred.
 
Additional Manufacturer Narrative
Manufacturers ref# (b)(4).Summary of investigational findings: during difficult intubation, the physician used a frova introducer with a double endotracheal tube.During fibroscopy a fragment of the blue introducer was found in the patient bronchus.Cook was unable to conduct a review of the device history record, as the lot number of the complaint device was not provided for the investigation.According to the instructions for use the 14.0 french frova introducer has been designed for placement of a single lumen endotracheal tube whose inner diameter is 6 mm or larger.Furthermore, the ifu warn that the tube and the introducer must be lubricated before use and that care must be taken when introducing/removing the introducer from the endotracheal tube, as contact with sharp edges on the internal surface of the endotracheal tube may cause small fragments to be shaved off the catheter introducer during introduction/removal.There are adequate controls in place to ensure that this type of device is manufactured to specifications.The device was not returned for evaluation.Based on the provided information the physician decided to use a double lumen endotracheal tube during the procedure.The ifu clearly specify that the frova device is intended for placement with use of a single lumen endotracheal tube with an inner diameter of 6mm or larger.Cook medical will continue to monitor for similar events.This report is required by the fda under 21cfr part 803.This report is based on unconfirmed information submitted by others.Neither the submission of this report nor any statement made in it is intended to be an admission that any cook device is defective or malfunctioned; that a death or serious injury occurred; or that any cook device caused or contributed to; or is likely to cause or contribute to a death or serious injury if a malfunction occurred.
 
Event or Problem Description
No additional information regarding the patient and/or event has been received since the previous medwatch report was sent.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FROVA INTUBATING INTRODUCER
Common Device Name
BTR TUBE, TRACHEAL (W/WO CONNECTOR)
Manufacturer (Section D)
WILLIAM COOK EUROPE
sandet 6
bjaeverskov 4632
DA  4632
Manufacturer (Section G)
WILLIAM COOK EUROPE
sandet 6, dk-4632
bjaeverskov
Manufacturer Contact
lissi walmann
sandet 6, dk-4632
bjaeverskov 
MDR Report Key16050126
Report Number3002808486-2022-01062
Device Sequence Number6200515
Product Code BTR
UDI-Device Identifier00827002483033
UDI-Public(01)00827002483033(17)251114(10)E4332947
Combination Product (Y/N)N
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Other,Foreign
Initial Reporter Occupation Pharmacist
Type of Report Initial,Followup,Followup
Report Date (Section B) 02/16/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date11/14/2025
Device Model NumberG48303
Device Catalogue NumberC-CAE-14.0-70-FIC
Device Lot NumberE4332947
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 12/21/2022
Supplement Date Received by Manufacturer02/10/2023
02/15/2023
Initial Report FDA Received Date12/27/2022
Supplement Report FDA Received Date02/14/2023
02/16/2023
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured11/14/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient SexUnknown
-
-