MAUDE Adverse Event Report: SMITH & NEPHEW, INC. LGN PS HIGH FLEX XLPE SZ 5-6 11MM; PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITH & NEPHEW, INC. LGN PS HIGH FLEX XLPE SZ 5-6 11MM; PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL Back to Search Results
Model Number 71453222
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Codes Pain (1994); Joint Laxity (4526)
Date of Event 11/27/2022
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Internal complaint reference: (b)(4).
 
Event or Problem Description
It was reported that, after a tka surgery was performed on (b)(6) 2012, the patient experienced instability and pain.This adverse event was treated by a revision surgery on (b)(6) 2022, in which a lgn ps high flex xlpe sz 5-6 11mm was exchanged.Patient's current health status is unknown.
 
Additional Manufacturer Narrative
H3, h6: the associated device was returned and evaluated.A visual inspection of the returned device revealed the device shows signs wear, likely prolonged use in the patient.However, with the information available, it can not be confirmed that this deformation was related to the reported instability and pain.A review of complaint history for the previous 12 months did not reveal similar events for the listed device.The clinical/medical evaluation concluded that per complaint details, a lgn ps hf insert revision was performed approximately 10 years post implantation due to instability and pain.As of the date of this medical investigation, no supporting clinical documentation has been provided.Reportedly, instability and pain necessitated the revision; however, definitive contributing clinical factors could not be concluded based on the limited information provided.The patient impact beyond the reported instability, pain and revision cannot be determined.No further medical assessment is warranted at this time.Device batch number was not provided, thus, an evaluation of the manufacturing records could not be performed.At this time, we do not have reason to suspect that the product failed to meet any product specifications at the time of manufacture.A review of the risk management and information for use files revealed this failure mode was previously identified.The anticipated risk level is still adequate.Assessment of historical escalated cases concluded that there are no prior actions related to this device and failure mode.With the results of this investigation the root cause of this event could not be determined.Factors that could contribute to the reported event include traumatic injury, limited range of motion, patient anatomy or abnormal loading of limb.Based on this investigation, the need for corrective action is not indicated.Should additional information be received, the complaint will be reopened.No further investigation is warranted for this complaint; however, we will continue to monitor future complaints and investigate as necessary.We consider this investigation closed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LGN PS HIGH FLEX XLPE SZ 5-6 11MM
Common Device Name
PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL
Manufacturer (Section D)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
Manufacturer (Section G)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
Manufacturer Contact
holly topping
7000 west william cannon drive
austin, TX 78735
5123913905
MDR Report Key16050247
Report Number1020279-2022-05083
Device Sequence Number6441904
Product Code JWH
UDI-Device Identifier00885556035351
UDI-Public00885556035351
Combination Product (Y/N)N
Initial Reporter StateIN
Initial Reporter CountryUS
PMA/510(K) Number
K071071
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2012
Device Explanted Year2022
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 01/19/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Model Number71453222
Device Catalogue Number71453222
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 11/30/2022
Supplement Date Received by Manufacturer01/18/2023
Initial Report FDA Received Date12/27/2022
Supplement Report FDA Received Date01/19/2023
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Hospitalization; Required Intervention;
Patient SexUnknown
-
-