This report is being supplemented to provide additional information based on the device evaluation and the legal manufacturer's final investigation.Please see the updates in sections d8, d9, h3, h4, h6, and h10.The device was returned and an evaluation was completed for it.During inspection, olympus confirmed the reported event, the image temporarily disappears.Additionally, the switching number two did not function and there was cable leakage; however, these defects are not considered severe enough to cause a potential adverse event.A review of the device history record found no deviations that could have caused or contributed to the reported issue.It has been over 9 years since the subject device was manufactured.Based on the results of the investigation, it is likely the image temporarily disappearing occurred due to a cable failure.However, a definitive root cause could not be determined.Olympus will continue to monitor field performance for this device.
|