Model Number 3545400
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Codes Breast Mass (2439); Breast Discomfort/Pain (4504)
Date of Event 12/05/2022
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Since the device has not been returned for analysis, no product failure analysis can be conducted, and no determination of possible contributing factors can be made.As such, the investigation will be closed.If the complaint device is received in the future, the investigation will be reopened and conducted as appropriate.A manufacturing record evaluation (mre) was performed, and no anomalies were found related to this complaint.In addition, the mre verifies that the device was manufactured in accordance with documented specification and procedures.Reason for device explant and/or reoperation: breast pain, breast mass.Mentor is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which mentor has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, mentor, or its employees that the report constitutes an admission that the device, mentor, or its employees caused or contributed to the potential event described in this report.Manufacturer¿s reference number: (b)(4).
 
Event or Problem Description
It was reported that a 36-year-old caucasian female who underwent a breast augmentation primary with a mentor memorygel, 400cc silicone breast prosthesis experienced bilateral, breast mass, breast pain and right-sided lymphadenopathy, and symptomatic rupture post procedure.The issue was diagnosed through ultrasound and mammogram examinations.The ultrasounds result stated that the multiple enlarged lymph nodes in right axillary tail are benign.Biopsy was cancelled due to benign nature of findings.As a result, patient underwent bilateral removal on (b)(6) 2022.This report relates to the left prosthesis.
 
Additional Manufacturer Narrative
Additional information received on january 04 2022 indicated that the patient underwent bilateral replacements as follow: right implant, ref:3505650bc, lot:7676015, sn:(b)(6), left implant, ref:3505650bc, lot: 9773701, sn:(b)(6) with capsulorrhaphy and partial capsulectomy on (b)(6) 20222.Per correct codification and clinician review of the new information, removed pc - lymphadenopathy and added pc ¿ granuloma to the right side.Manufacturer¿s reference number: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MENTOR MEMORYGEL BREAST IMPLANT
Common Device Name
PROSTHESIS, BREAST, NONINFLATABLE, INTERNAL, SILICONE GEL-FILLED
Manufacturer (Section D)
MENTOR TEXAS
3041 skyway circle north
irving TX 75038
Manufacturer (Section G)
MENTOR TEXAS
3041 skyway circle north
irving TX 75038
Manufacturer Contact
kate karberg
3041 skyway circle north
irving, TX 75038
3035526892
MDR Report Key16051144
Report Number1645337-2022-15457
Device Sequence Number6209967
Product Code FTR
UDI-Device Identifier00081317007142
UDI-Public00081317007142
Combination Product (Y/N)N
Initial Reporter StateTX
Initial Reporter CountryUS
PMA/510(K) Number
P030053
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2018
Device Explanted Year2022
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Other
Type of Report Initial,Followup
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date06/21/2022
Device Model Number3545400
Device Catalogue Number3545400
Device Lot Number7480987
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 12/06/2022
Supplement Date Received by Manufacturer01/04/2023
Initial Report FDA Received Date12/27/2022
Supplement Report FDA Received Date01/12/2023
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured06/22/2017
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age36 YR
Patient SexFemale
Patient EthnicityNon Hispanic
Patient RaceWhite
-
-