MEDTRONIC PUERTO RICO OPERATIONS CO. 640G INSULIN PUMP MMT-1712K; PUMP, INFUSION, INSULIN, TO BE USED WITH INVASIVE GLUCOSE SENSOR
|
Back to Search Results |
|
Model Number MMT-1712K |
Device Problems
Break (1069); Device Difficult to Program or Calibrate (1496); Visual Prompts will not Clear (2281)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 12/13/2022 |
Event Type
malfunction
|
Event Description
|
Information received by medtronic indicated that the insulin pump fall from height, buttons were not responding and the screen got white.Troubleshooting was performed and the issue was not resolved.No harm requiring medical intervention was reported.Customer was calling back after installing a new battery and monitoring the insulin pump for two hours and frozen display recurred.The customer will discontinue using the device and was advised to revert to the backup plan as per hcp instructions.The pump will be returned for analysis.
|
|
Manufacturer Narrative
|
(b)(4).Currently it is unknown whether or not the device may have caused or contributed to the event as no product has been returned.The device will be returned for analysis and further information will follow once the analysis has been completed.No conclusion can be drawn at this time.The insulin pump involved in this event is the ngp 640g insulin pump which is not marketed in the united states.However, the device is similar to the ngp insulin pump, which is marketed in the united states.Medtronic, inc.(medtronic) is submitting this report to comply with 21 c.F.R.Part 803, the medical device reporting regulation.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information in the time allotted and has provided as much information as is available to the company as of the submission date this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employees caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any "defects" or has "malfunctioned".These words are included in the fda 3500a form and are fixed items for selection created by the fda, to categorize the type of event solely for the purpose of reporting pursuant to part 803.Medtronic objects to the use of these words and others like it because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.
|
|
Manufacturer Narrative
|
The pump passed the displacement test, sleep current test, active current test, and self test.All buttons function properly.Successfully downloaded history files and traces using thus.The power management graph confirmed the unloaded voltage (ul vlith) and loaded voltage (loaded vlith) were within spec range.Pump was cut open to perform visual inspection and found moisture damage to the electronic assembly, and motor.The j1 connector on pcba 1 was locked properly during visual inspection.Test p-cap and reservoir locked properly into reservoir compartment during testing.The following were noted during visual inspection: scratched case, cracked case (battery tube), pillowing keypad overlay, corroded home switch.Keypad buttons functioned properly during analysis and passed all required testing.Unresponsive keypad was not confirmed.Frozen screen not confirmed.Unspecified cosmetic damage confirmed.Medtronic, inc.(medtronic) is submitting this report to comply with 21 c.F.R.Part 803, the medical device reporting regulation.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information in the time allotted and has provided as much information as is available to the company as of the submission date this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employees caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any "defects" or has "malfunctioned".These words are included in the fda 3500a form and are fixed items for selection created by the fda, to categorize the type of event solely for the purpose of reporting pursuant to part 803.Medtronic objects to the use of these words and others like it because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.
|
|