• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BECTON, DICKINSON & CO. (BROKEN BOW) BD VACUTAINER® K2 EDTA (K2E) 3.6MG BLOOD COLLECTION TUBES; BLOOD SPECIMEN COLLECTION DEVICE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BECTON, DICKINSON & CO. (BROKEN BOW) BD VACUTAINER® K2 EDTA (K2E) 3.6MG BLOOD COLLECTION TUBES; BLOOD SPECIMEN COLLECTION DEVICE Back to Search Results
Model Number 367841
Device Problem Short Fill (1575)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 11/23/2022
Event Type  malfunction  
Event Description
It was reported when using the bd vacutainer® k2 edta (k2e) 3.6mg blood collection tubes had insufficient suction during use.The following information was provided by the initial reporter.The customer stated: "during the blood drawing process, the nurse found that after sucking out about 1ml of blood with the negative pressure suction tube of batch number 2018841, the negative pressure in the tube was insufficient to continue sucking out blood, and this happened in all boxes.".
 
Manufacturer Narrative
Investigation summary: "no customer samples and no photos were received in support of this complaint from catalog 367841, lot number 2018841.Therefore, 100 retentions were visually inspected with the hemogard closure assembly correctly assembled and placed on the tubes.There were no cocked stoppers identified that would cause the hemogard closure assembly to not be applied correctly.The quantity of blood drawn into evaluated tubes varies with altitude, ambient temperature, barometric pressure, tube age, venous pressure, and filling technique.10 production lot in-house retention tubes were inspected with 0 visible defects.A draw test was performed at the manufacturing site on the 10-production lot in-house retention tubes.All tubes were within specification limits.Bd was unable to confirm the customer¿s indicated failure modes of underfill based on the investigation completed.The defect was not observed in the retention sample testing.".
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BD VACUTAINER® K2 EDTA (K2E) 3.6MG BLOOD COLLECTION TUBES
Type of Device
BLOOD SPECIMEN COLLECTION DEVICE
Manufacturer (Section D)
BECTON, DICKINSON & CO. (BROKEN BOW)
150 south 1st avenue
broken bow NE 68822
Manufacturer (Section G)
BECTON, DICKINSON & CO. (BROKEN BOW)
150 south 1st avenue
broken bow NE 68822
Manufacturer Contact
phillip emmert
9450 south state street
sandy, UT 84070
8015296192
MDR Report Key16051895
MDR Text Key306840580
Report Number1917413-2022-00809
Device Sequence Number1
Product Code JKA
Combination Product (y/n)N
Reporter Country CodeCH
PMA/PMN Number
BK050036
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,User Facility
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 12/15/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/27/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Expiration Date04/30/2023
Device Model Number367841
Device Catalogue Number367841
Device Lot Number2018841
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received11/30/2022
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured01/18/2022
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-