Model Number U411
Medical Device Problem Code Delayed Program or Algorithm Execution (4034)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 11/28/2022
Type of Reportable Event Malfunction
Event or Problem Description
The initial reporter alleged there was a mismatch of patient data for an unspecified number of urine patient samples tested on the cobas u 411 urine analyzer.The software version of the analyzer is 3.3.3.0019.On (b)(6) 2022, the reporter observed that the results obtained did not match the id numbers.They then processed sample by sample.On (b)(6) 2022, the issue persisted.The reporter noted that when two samples were processed, two results come out but the first comes out with a different id in the indicated free sequence and the second comes out with the id of the sample that was entered first.The reporter used a worklist.
 
Additional Manufacturer Narrative
The investigation determined that the issue cannot be confirmed based on the data submitted covering the date of the event.The customer reported using a worklist.Product labeling states "the test strip is allocated to the first sample on the list." it was confirmed that the analyzer is working correctly.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
COBAS U 411 URINE ANALYZER
Common Device Name
AUTOMATED URINALYSIS SYSTEM
Manufacturer (Section D)
ROCHE DIAGNOSTICS
9115 hague road
indianapolis IN 46250 0457
Manufacturer (Section G)
ROCHE DIAGNOSTICS GMBH
sandhoferstrasse 116
na
mannheim (baden-wurttemberg) 68305
GM   68305
Manufacturer Contact
amy nelson
9115 hague road
na
indianapolis, IN 46250
3174767531
MDR Report Key16052369
Report Number1823260-2022-04184
Device Sequence Number20419592
Product Code KQO
UDI-Device Identifier04015630925636
UDI-Public04015630925636
Combination Product (Y/N)N
Initial Reporter CountryMX
PMA/510(K) Number
K093555
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,User Facility
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date (Section B) 12/27/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberU411
Device Catalogue Number04906969001
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 11/29/2022
Initial Report FDA Received Date12/27/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient SexUnknown
-
-