| Model Number |
U411 |
| Medical Device Problem Code |
Delayed Program or Algorithm Execution (4034)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
11/28/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
The initial reporter alleged there was a mismatch of patient data for an unspecified number of urine patient samples tested on the cobas u 411 urine analyzer.The software version of the analyzer is 3.3.3.0019.On (b)(6) 2022, the reporter observed that the results obtained did not match the id numbers.They then processed sample by sample.On (b)(6) 2022, the issue persisted.The reporter noted that when two samples were processed, two results come out but the first comes out with a different id in the indicated free sequence and the second comes out with the id of the sample that was entered first.The reporter used a worklist.
|
| |
|
Additional Manufacturer Narrative
|
|
The investigation determined that the issue cannot be confirmed based on the data submitted covering the date of the event.The customer reported using a worklist.Product labeling states "the test strip is allocated to the first sample on the list." it was confirmed that the analyzer is working correctly.
|
| |
|