• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ALCON RESEARCH, LLC - HUNTINGTON ACRYSOF SINGLEPIECE IOL; INTRAOCULAR LENS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ALCON RESEARCH, LLC - HUNTINGTON ACRYSOF SINGLEPIECE IOL; INTRAOCULAR LENS Back to Search Results
Model Number SA60AT
Medical Device Problem Code Scratched Material (3020)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 11/25/2022
Type of Reportable Event Malfunction
Event or Problem Description
A nurse reported that during the cataract surgery with intraocular lens (iol) implant procedure optic found to be scratched after implantation in the eye.The surgeon explanted the lens with new one.The surgery was completed without any patient adverse events.Additional information has been requested.
 
Additional Manufacturer Narrative
A sample device was not returned for analysis.Product history records were reviewed and documentation indicated the product met release criteria.Root cause: root cause has not been identified.There have been no other complaints reported in the lot.The manufacturer internal reference number is: (b)(4).
 
Additional Manufacturer Narrative
The lens was returned for evaluation.Solution is dried on the lens.The lens was returned in multiple pieces.Scratches are observed on the lens.A deep gouge is observed on the optic portion of one of the pieces.The file indicates the use of a qualified cartridge with an unspecified handpiece.Two viscoelastics were indicated, only one is qualified for this lens/cartridge combination.Lens damage was observed.All lenses are 100% inspected for cosmetic attributes and the damage exhibited by the returned complaint sample would not have met company release criteria.Based on our observation, it can be reasonably concluded that the damage is not manufacturing related.Due to the presence of surgical solution and the condition of the returned sample, the damage is most likely related to customer handling.The manufacturer internal reference number is: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ACRYSOF SINGLEPIECE IOL
Common Device Name
INTRAOCULAR LENS
Manufacturer (Section D)
ALCON RESEARCH, LLC - HUNTINGTON
6065 kyle lane
huntington WV 25702
Manufacturer (Section G)
ALCON RESEARCH, LLC - HUNTINGTON
6065 kyle lane
huntington WV 25702
Manufacturer Contact
jonathan schlech
6201 south freeway
mail stop ab2-6
fort worth, TX 76134
8007579780
MDR Report Key16052673
Report Number1119421-2022-02741
Device Sequence Number6442459
Product Code HQL
UDI-Device Identifier00380652555784
UDI-Public00380652555784
Combination Product (Y/N)N
Initial Reporter CountrySN
PMA/510(K) Number
P930014
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Nurse
Type of Report Initial,Followup
Report Date (Section B) 03/13/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberSA60AT
Device Catalogue NumberSA60AT.190
Device Lot Number15441211
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 12/02/2022
Supplement Date Received by Manufacturer02/13/2023
Initial Report FDA Received Date12/27/2022
Supplement Report FDA Received Date03/13/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured06/29/2022
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
CUSTOM-PAK; DUOVISC; MONARCH CARTRIDGE C; MONARCH INJECTOR, UNSPECIFIED
Patient SexUnknown
-
-