ALCON RESEARCH, LLC - HUNTINGTON ACRYSOF IQ VIVITY EXTENDED VISION IOL; INTRAOCULAR LENS Back to Search Results
Model Number DFT015
Medical Device Problem Code Scratched Material (3020)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 11/18/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
A sample device was not returned for analysis.Complaint history and product history records were reviewed and documentation indicated the product met release criteria.Root cause has not been identified.There have been no other complaints reported in the lot number.The manufacturer internal reference number is: (b)(4).
 
Event or Problem Description
A physician reported that during an intraocular lens (iol) implant procedure, there was immediate post implantation paracentral scratches.There was no patient harm.No further information available.
 
Additional Manufacturer Narrative
A qualified cartridge and handpiece were indicated with a non-qualified viscoelastic.The root cause for the reported lens damage may be related to a failure to follow the instructions for use (ifu).It was stated that viscoelastic was used so that the window with the image of the iol on the cartridge was full.So the first 5-6mm seen from the loading side.This would only be at the loading area.This is not per the ifu.The ifu instructs to completely fill the cartridge with ovd immediately prior to loading and delivery of the lens.Do not attempt to load the lens without adequate ovd in the device.Not adequately filling the device with viscoelastic will result in inadequate coverage of lens and the lens fold path with ovd, which may result in damage.Company foldable iols are qualified for use with an company qualified delivery system (handpiece and cartridge) and ophthalmic viscosurgical device (ovd) combination.The use of an unqualified combination may cause damage to the iol and potential complications during the implantation process.The manufacturer internal reference number is: (b)(4).
 
Event or Problem Description
Additional information has been requested and received stating that patients were not affected and the problem was also solved.It was further stated that the small micro-scratches do not affect the patient's vision at all.Finally it was found out if the cartridge was filled properly the micro-scratches no longer occur.The problem has not been present for months.This was a mini problem which was solved and most importantly this problem has never made any clinical effects/problems for patients.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ACRYSOF IQ VIVITY EXTENDED VISION IOL
Common Device Name
INTRAOCULAR LENS
Manufacturer (Section D)
ALCON RESEARCH, LLC - HUNTINGTON
6065 kyle lane
huntington WV 25702
Manufacturer (Section G)
ALCON RESEARCH, LLC - HUNTINGTON
6065 kyle lane
huntington WV 25702
Manufacturer Contact
jonathan schlech
6201 south freeway
mail stop ab2-6
fort worth, TX 76134
8007579780
MDR Report Key16052677
Report Number1119421-2022-02740
Device Sequence Number17752929
Product Code POE
UDI-Device Identifier00380652421317
UDI-Public00380652421317
Combination Product (Y/N)N
Initial Reporter CountryGM
PMA/510(K) Number
P930014
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 02/16/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberDFT015
Device Lot Number15332387
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 11/30/2022
Supplement Date Received by Manufacturer01/18/2023
Initial Report FDA Received Date12/27/2022
Supplement Report FDA Received Date02/16/2023
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured01/07/2022
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
1.2% OF ZEISS; MONARCH II LOADING FORCEPS; MONARCH III IOL DELIVERY SYSTEM, CARTRIDGE C; MONARCH III IOL DELIVERY SYSTEM, INJECTOR