• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ROCHE DIAGNOSTICS COBAS E 801 ANALYTICAL UNIT; IMMUNOCHEMISTRY ANALYZER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ROCHE DIAGNOSTICS COBAS E 801 ANALYTICAL UNIT; IMMUNOCHEMISTRY ANALYZER Back to Search Results
Model Number E801
Medical Device Problem Codes High Test Results (2457); Non Reproducible Results (4029)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 12/06/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Qc was acceptable on the day of the event.The alarm trace showed several sample clot, sample short, and sample foam detection errors between (b)(6) and (b)(6) 2022.Based on the available data, a general reagent issue could be excluded.The investigation did not identify a product problem.The root cause of the event could not be determined.
 
Event or Problem Description
There was an allegation of questionable elecsys vitamin d assay results for 1 patient sample on a cobas e 801 analytical unit.On (b)(6) 2022, the initial vitamin d result was 100 ng/ml with flag.On (b)(6) 2022, the sample was repeated twice and the results were 11.2 ng/ml and 20.8 ng/ml.The initial result was not reported outside of the laboratory.The reagent lot is 601999 and the expiration date is 28-feb-2023.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
COBAS E 801 ANALYTICAL UNIT
Common Device Name
IMMUNOCHEMISTRY ANALYZER
Manufacturer (Section D)
ROCHE DIAGNOSTICS
9115 hague road
indianapolis IN 46250 0457
Manufacturer (Section G)
HITACHI HIGH TECH CORP.
882 ichige hitachinaka
na
ibaraki 312-8 504
JA   312-8504
Manufacturer Contact
amy nelson
9115 hague road
na
indianapolis, IN 46250
3174767531
MDR Report Key16052740
Report Number1823260-2022-04188
Device Sequence Number12326702
Product Code JJE
UDI-Device Identifier04015630940127
UDI-Public04015630940127
Combination Product (Y/N)N
Initial Reporter CountryTH
PMA/510(K) Number
K060373
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,User Facility
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date (Section B) 12/27/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberE801
Device Catalogue Number08454345001
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 12/07/2022
Initial Report FDA Received Date12/27/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient SexUnknown
-
-