| Model Number |
E801 |
| Medical Device Problem Codes |
High Test Results (2457); Non Reproducible Results (4029)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
12/06/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Qc was acceptable on the day of the event.The alarm trace showed several sample clot, sample short, and sample foam detection errors between (b)(6) and (b)(6) 2022.Based on the available data, a general reagent issue could be excluded.The investigation did not identify a product problem.The root cause of the event could not be determined.
|
| |
|
Event or Problem Description
|
|
There was an allegation of questionable elecsys vitamin d assay results for 1 patient sample on a cobas e 801 analytical unit.On (b)(6) 2022, the initial vitamin d result was 100 ng/ml with flag.On (b)(6) 2022, the sample was repeated twice and the results were 11.2 ng/ml and 20.8 ng/ml.The initial result was not reported outside of the laboratory.The reagent lot is 601999 and the expiration date is 28-feb-2023.
|
| |
|