• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: OLYMPUS WINTER & IBE GMBH TELESCOPE "OES ELITE", 4 MM, 0°, HD, AUTOCLAVABLE; HD AUTOCLAVABLE TELESCOPE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

OLYMPUS WINTER & IBE GMBH TELESCOPE "OES ELITE", 4 MM, 0°, HD, AUTOCLAVABLE; HD AUTOCLAVABLE TELESCOPE Back to Search Results
Model Number WA2T400A
Medical Device Problem Code Crack (1135)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 11/28/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The repair inspection confirmed the customer¿s reported defect, blurry with shadow on the image due to broken lens in the optical system.The inspection identified a broken/cracked eyepiece funnel (reportable).Additionally, the inspection noted, the rigid outer tube is bent, there is distal end fiber damage, and minor fading of the markings on the model ring.The investigation is ongoing.However, if additional information becomes available, this report will be supplemented accordingly.In general, the end-user is required to inspect the device for defects, check the function of all devices and have alternate equipment prior to use.
 
Event or Problem Description
Olympus was informed that the customer ¿oes elite¿ high-definition autoclavable telescope was returned to the olympus repair center for a reported ¿blurry lens¿.The customer reported defect was found at reprocessing.No death injury or infection and no person or other was affected has been reported to olympus.However, during the repair inspection, olympus found the eyepiece funnel is cracked.This report is being submitted to capture the broken eye piece component.
 
Additional Manufacturer Narrative
This report is being supplemented to provide additional information based on the legal manufacturer's final investigation.Please see the updates in sections h6 and h10.A review of the device history record found no deviations that could have caused or contributed to the reported issue.It has been over 6 years since the subject device was manufactured.Based on the results of the investigation, it is likely the reported issue occurred due to excessive use as well as wear and tear.However, a definitive root cause could not be determined.Olympus will continue to monitor field performance for this device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TELESCOPE "OES ELITE", 4 MM, 0°, HD, AUTOCLAVABLE
Common Device Name
HD AUTOCLAVABLE TELESCOPE
Manufacturer (Section D)
OLYMPUS WINTER & IBE GMBH
kuehnstrasse 61
hamburg, ds 22045
GM  22045
Manufacturer (Section G)
OLYMPUS WINTER & IBE GMBH
kuehnstrasse 61
hamburg
Manufacturer Contact
todd brill
800 west park drive
westborough, MA 01581
5082077661
MDR Report Key16053115
Report Number9610773-2022-00744
Device Sequence Number6220070
Product Code FGC
UDI-Device Identifier04042761077491
UDI-Public04042761077491
Combination Product (Y/N)N
Initial Reporter StateCA
Initial Reporter CountryUS
PMA/510(K) Number
K151191
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,User Facility,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 03/16/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberWA2T400A
Device Catalogue NumberWA2T400A
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/30/2022
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 11/28/2022
Supplement Date Received by Manufacturer02/21/2023
Initial Report FDA Received Date12/27/2022
Supplement Report FDA Received Date03/16/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured06/16/2016
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Patient SexUnknown
-
-