| Model Number |
E801 |
| Medical Device Problem Codes |
False Positive Result (1227); Non Reproducible Results (4029)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
12/06/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Calibration and qc were acceptable prior to the event.Based on the available data, a general reagent issue could be excluded.The investigation did not identify a product problem.The root cause of the event could not be determined.
|
| |
|
Event or Problem Description
|
|
There was an allegation of questionable elecsys anti-hav assay gen.2 results for 1 patient sample on a cobas e 801 analytical unit.The initial anti-hav result was 0.790 coi, interpreted as reactive.The sample was recentrifuged and repeated.The repeat result was 1.19 coi, interpreted as non-reactive.The next day, the sample was recentrifuged again and repeated again.The second repeat result was 1.19 coi, interpreted as non-reactive.The initial result was not reported outside of the laboratory.The reagent lot is 625717 and the expiration date is 31-jul-2023.
|
| |
|