• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEPUY MITEK LLC US 30° SUTURE GRASPER; SUTURE/NEEDLE PASSER, REUSABLE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY MITEK LLC US 30° SUTURE GRASPER; SUTURE/NEEDLE PASSER, REUSABLE Back to Search Results
Model Number 251721
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code Foreign Body In Patient (2687)
Date of Event 12/01/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
This report is being submitted in pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by depuy mitek or its employees that the report constitutes an admission that the device, depuy mitek, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.Udi: (b)(4).Patient identifier: (b)(6).To date the device has not been returned.If the device or further details are received at a later date a supplemental medwatch will be sent.
 
Event or Problem Description
This is report 1 of 2 for (b)(4).It was reported by a healthcare professional in singapore that during an unknown procedure on (b)(6) 2022, it was observed that the wire tip on the 30° suture grasper device broke while the surgeon was trying to extract the handle out.According to the report, the device was used to penetrate through labral tissue.Changed to another one but the same issue happened again.It was reported that the surgeon tried to scope around to look for the missing 3mm wire but unable to locate it; and therefore, the surgeon ordered for a post op x-ray to be conducted.It was reported that the patient was in stable condition.There were no adverse patient consequences reported.No additional information was provided.
 
Additional Manufacturer Narrative
This report is being submitted in pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by depuy mitek or its employees that the report constitutes an admission that the device, depuy mitek, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.Investigation summary: the complaint device is not being returned, therefore unavailable for a physical evaluation.However, photos were provided.Upon visual inspection of the photos, it appeared that tip was broken.Based on the condition, this complaint reported was confirmed.A manufacturing record evaluation was performed for the finished device 8l06404, and no non-conformances were identified.The photo does not provide enough evidence to determine root cause.Physical evaluation should provide more information to discern a possible root cause.A definite root case cannot be determined for this failure.The damages are associated with levering during insertion or excess tension applied on inserter which caused damage.However, it cannot be conclusively affirmed.As per ifu, excessive force while using this instrument for tissue approximation is not recommended and may result in damaging the device.Further, a review into the depuy synthes mitek complaints system revealed no other complaints of any type for this lot that was released to distribution.At this point in time, no corrective action is required, and no further action is warranted.However, depuy synthes mitek will continue to track any related complaints within this device family as a means of monitoring the extent with which this complaint is observed in the field.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
30° SUTURE GRASPER
Common Device Name
SUTURE/NEEDLE PASSER, REUSABLE
Manufacturer (Section D)
DEPUY MITEK LLC US
325 paramount drive
raynham MA 02767
Manufacturer (Section G)
DEPUY MITEK LLC US
325 paramount drive
raynham MA 02767
Manufacturer Contact
kate karberg
325 paramount drive
raynham, MA 02767
3035526892
MDR Report Key16055341
Report Number1221934-2022-04238
Device Sequence Number17751639
Product Code NBH
UDI-Device Identifier10886705012299
UDI-Public10886705012299
Combination Product (Y/N)N
Initial Reporter CountrySN
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,User Facility,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date03/31/2026
Device Model Number251721
Device Catalogue Number251721
Device Lot Number8L06404
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 12/05/2022
Supplement Date Received by Manufacturer01/02/2023
Initial Report FDA Received Date12/27/2022
Supplement Report FDA Received Date01/03/2023
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured04/30/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient Age34 YR
Patient SexMale
-
-