Follow-up report #1 is to provide fda with corrected data and new informations.New information: b4, b5, g6, and h2.Corrected data: -g3 (date received by manufacturer): november 24, 2022.-e1 (initial reporter name and address).
|
|
Follow-up report #2 is to provide fda with the results of the device investigation.New information: b4, b5, d9, g3, g6, h2, h3, h6 (type of investigation, investigation findings, investigation conclusions), h7, and h10.The defective inner part of the forceps was sent to richard wolf in knittlingen for examination.It was found that one fork on the holder part was broken.The second fork was bent and cracks were already visible.The holder part was torn off.The two jaws were not available for examination.However, it can be seen on a picture that all parts of the inner part of the forceps were present after the recovery of the broken pieces.The bowel grasping forceps 8393249 from lot 4500250452 was booked into stock at richard wolf in knittlingen with a quantity of (b)(4) pieces on 04/12/2018.The lot 4500250452 was completely sold out in the period 13/12/2018 to 22/02/2019.To the company hsl (distributor in ireland) 2 grasping forceps from lot 4500250452 were sold on 22.02.2019.On 07/02/2019 a complaint was made about a grasping forceps from lot 4500250452 from oman.The reason given for the complaint was "tissues gets strip while holds in process".However, the inspection did not reveal any defect.No other intestinal forceps from lot 45002580452 have been complained about so far.In the corresponding risk analysis b1-2 rev.05, manufacturing-related, handling-related and design-related hazards with regard to a functional impairment as well as risks due to an unusable product with the corresponding extent of damage and the assumed probability of occurrence were considered and assessed with an acceptable risk.Since no new risks have arisen from the investigation of the current complaint case, the risk assessment remains valid in view of the facts described.The user is advised in the associated instructions for use ga-s027 / en / 2019-11 v10.0 / pk19-9850 under chapter 10 application that the limited stability of the product must be taken into account during application."10 use warning.The products have only limited strength! excessive force will cause damage, impair the function and therefore endanger the patient.Immediately before and after use, check the products for damage, loose parts and completeness.Make sure that no missing instrument parts remain in the patient.Do not use products that are damaged, incomplete or have loose parts.".Visual inspections before and after use as described in chapters 10 and 11.2 serve to detect damage at an early stage.The exact circumstances that led to the loss of parts not being noticed during the actual operation are not known.The patient was examined radiographically and the broken-off parts of the intestinal forceps were completely removed from the patient.Due to the strong mechanical deformation of the instrument, mechanical overload is considered to be the cause.Richard wolf assumes user error.A general product or construction problem is not recognisable.From a manufacturing or design point of view, no systematic defect could be deduced at this time, so that no measures were considered necessary.
|