• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BECTON, DICKINSON & CO. (SPARKS) BD BACTEC¿ FX, INSTRUMENT TOP, PACKAGED; SYSTEM, BLOOD CULTURING

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BECTON, DICKINSON & CO. (SPARKS) BD BACTEC¿ FX, INSTRUMENT TOP, PACKAGED; SYSTEM, BLOOD CULTURING Back to Search Results
Model Number 441386
Medical Device Problem Code Unintended Movement (3026)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 12/15/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Initial reporter phone: (b)(6).A device evaluation and/or device history review is anticipated but is not complete.Upon completion, a supplemental report will be filed.
 
Event or Problem Description
It was reported that bd bactec¿ fx, instrument top, packaged right front leveling foot dislodged from strut.The following information was provided by the initial reporter: front right leveling foot broken off of strut.No injuries were reported.
 
Additional Manufacturer Narrative
H.6 investigation summary: customer contacted bd support regarding their bactec fx bottom 441386 sn fb1414 alleging the leveling foot had broken.A bd field service engineer went onsite to investigate and confirmed the issue.The fse replaced the foot and ensured the instrument was functioning properly before turning back over to the customer.This is a confirmed complaint.Dhr review is not required as this complaint has been confirmed as a failure of the bd product through other elements as noted in the investigation.Service history review shows no prior complaints have been opened related to this failure mode.Quality did not receive samples for this complaint and therefore returned sample analysis could not occur.The root cause is unknown.Review of risk management files confirms there are no new or modified risks associated with this failure mode.Bd quality will continue to monitor for trends associated with the failure mode described in this complaint.H3 other text : see h.10.
 
Event or Problem Description
It was reported that bd bactec¿ fx, instrument top, packaged right front leveling foot dislodged from strut.The following information was provided by the initial reporter: front right leveling foot broken off of strut.No injuries were reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BD BACTEC¿ FX, INSTRUMENT TOP, PACKAGED
Common Device Name
SYSTEM, BLOOD CULTURING
Manufacturer (Section D)
BECTON, DICKINSON & CO. (SPARKS)
7 loveton circle
sparks MD 21152
Manufacturer (Section G)
BECTON, DICKINSON & CO. (SPARKS)
7 loveton circle
sparks MD 21152
Manufacturer Contact
phillip emmert
9450 south state street
sandy, UT 84070
8015296192
MDR Report Key16056776
Report Number1119779-2022-01548
Device Sequence Number14676583
Product Code MDB
UDI-Device Identifier00382904413866
UDI-Public00382904413866
Combination Product (Y/N)N
Initial Reporter CountryCA
PMA/510(K) Number
K915796
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Other,User Facility
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 08/15/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Other
Device Model Number441386
Device Catalogue Number441386
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 12/15/2022
Supplement Date Received by Manufacturer08/15/2023
Initial Report FDA Received Date12/27/2022
Supplement Report FDA Received Date08/17/2023
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured03/27/2014
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-