LAKE REGION MEDICAL THRUWAY GUIDEWIRE; WIRE, GUIDE, CATHETER Back to Search Results
Model Number M001492971
Medical Device Problem Code Entrapment of Device (1212)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 12/06/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Product was not received at the time of this report; therefore, no physical or visual analysis of the product could be performed.A review of the device history records of the specimen device does not present any indication of manufacturing defect or anomaly that could have impacted on the event as reported.During manufacturing and packaging of this product the operators are instructed to 100% visually inspect for any obvious defect prior to shipment.The history records indicate this product was final inspection tested at lake region medical and was determined to be acceptable.As noted in the device instructions for use (dfu), "free movement of the guidewire within the catheter is an important feature of a steerable guidewire system because it gives the user valuable tactile information.Test the system for any resistance prior to use." at this time, it is not possible to assign a definitive root cause for the event as reported.If there is any further relevant information received, a follow up medwatch report will be submitted.
 
Event or Problem Description
Event description: per cnf, it was reported that: event others (please specify the details in the event description column.) shaping? no.The wire and catheter got stuck while using jetstream xc 2.1/3.0lot29360600 so it was completely removed.Procedure outcome: completed with another of same device introducer sheath used? yes.Able to use the device guidewire used? unknown.Guide catheter used? yes.Able to use the device balloon catheter used? yes.Able to use the device stent used? unknown.Inflation device used? yes.Able to use the device imaging catheter? unknown.
 
Additional Manufacturer Narrative
Part a patient information was updated to reflect the additional information received on january 5, 2023.
 
Additional Manufacturer Narrative
Device evaluation: a review of the device history records of the specimen device does not present any indication of manufacturing defect or anomaly that could have impacted on the event as reported.During manufacturing and packaging of this product the operators are instructed to 100% visually inspect for any obvious defect prior to shipment.The history records indicate this product was final inspection tested at lake region medical and was determined to be acceptable.The specimen presented extensive kink/bend damage of varying severity and frequency scattered over the length of the device.The damage presented appeared consistent with manipulation of the device against resistance.The event description did not make any mention of kink/bend damage; as such, it could have been incurred during post-event handling to separate the thruway wire from the jetstream.Our investigation was unable to confirm that the product did not meet specification prior to shipment.The investigation concluded that the product met specification at the time of shipment.At this time, it is not possible to assign a definitive root cause for the event as reported.As noted in the device instructions for use (dfu), free movement of the guidewire within the catheter is an important feature of a steerable guidewire system because it gives the user valuable tactile information.Test the system for any resistance prior to use.At this time, it is not possible to assign a definitive root cause for the event as reported.Based on the information provided, it appeared that clinical and/or procedural factors have contributed to the event as reported.If any further relevant information is received, a follow up medwatch report will be submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
THRUWAY GUIDEWIRE
Common Device Name
WIRE, GUIDE, CATHETER
Manufacturer (Section D)
LAKE REGION MEDICAL
340 lake hazeltine drive
chaska MN 55318
Manufacturer (Section G)
LAKE REGION MEDICAL
340 lake hazeltine drive
chaska MN 55318
Manufacturer Contact
sharon seifert
340 lake hazeltine drive
chaska, MN 55318
9526418518
MDR Report Key16056836
Report Number2126666-2022-00052
Device Sequence Number4937164
Product Code DQX
UDI-Device Identifier08714729717188
UDI-Public08714729717188
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K042338
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Distributor
Initial Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date (Section B) 02/06/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date07/31/2025
Device Model NumberM001492971
Device Catalogue Number39525
Device Lot Number7063259
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 12/15/2022
Supplement Date Received by Manufacturer01/05/2023
02/01/2023
Initial Report FDA Received Date12/27/2022
Supplement Report FDA Received Date01/11/2023
02/06/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured07/14/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Patient Age83 YR
Patient SexFemale
Patient Weight41 KG
-
-