• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ABBOTT VASCULAR NC TREK CORONARY DILATATION CATHETER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ABBOTT VASCULAR NC TREK CORONARY DILATATION CATHETER Back to Search Results
Model Number 1012449-15
Medical Device Problem Codes Material Rupture (1546); Difficult to Advance (2920)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 11/30/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The device will not be returned for evaluation, the device was reportedly discarded.Investigation is not yet complete.A follow-up report will be submitted with all additional relevant information.
 
Event or Problem Description
It was reported that the procedure was to treat a lesion in the mid and distal left anterior descending artery (lad) with 100% stenosis, heavy calcification and heavy tortuosity.The 3.0x15mm nc trek balloon dilatation catheter (bdc) was soaked in saline and prepared (air aspiration) outside the anatomy prior to use.There was no resistance when the protective sheath was removed.There was resistance with the lesion during advancement and was attempted to be inflated but the balloon ruptured after the first inflation at 18 atmospheres (atms) for 5 seconds.There was no resistance during removal.A non-abbott balloon was used to complete the procedure.There was no adverse patient effect and there was no clinically significant delay in the procedure.No additional information was provided.
 
Event or Problem Description
It was reported that the procedure was to treat a lesion in the mid and distal left anterior descending artery (lad) with 100% stenosis, heavy calcification and heavy tortuosity.The 3.0x15mm nc trek balloon dilatation catheter (bdc) was soaked in saline and prepared (air aspiration) outside the anatomy prior to use.There was no resistance when the protective sheath was removed.There was resistance with the lesion during advancement and was attempted to be inflated but the balloon ruptured after the first inflation at 18 atmospheres (atms) for 5 seconds.There was no resistance during removal.A non-abbott balloon was used to complete the procedure.There was no adverse patient effect and there was no clinically significant delay in the procedure.No additional information was provided.
 
Additional Manufacturer Narrative
The device was not returned for evaluation.A review of the lot history record identified no manufacturing nonconformities issued to the reported lot that would have contributed to this event.Additionally, a review of the complaint history identified no other similar incidents and/or complaints from this lot.The investigation determined the reported difficulties appear to be related to circumstances of the procedure.There is no indication of a product quality issue with respects to the design, manufacture, or labeling of the device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
NC TREK CORONARY DILATATION CATHETER
Common Device Name
CORONARY DILATATION CATHETER
Manufacturer (Section D)
ABBOTT VASCULAR
26531 ynez rd.
temecula CA 92591 4628
Manufacturer (Section G)
ABBOTT VASCULAR COSTA RICA, REG # 3009564766
52 calle 3 b31 coyol free zone
el coyol alajuela
CS  
Manufacturer Contact
lindsey bell
26531 ynez rd.
temecula, CA 92591-4628
9519143996
MDR Report Key16056879
Report Number2024168-2022-13143
Device Sequence Number12088935
Product Code LOX
UDI-Device Identifier08717648151873
UDI-Public08717648151873
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K103153
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative,Distributor
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 01/18/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date08/31/2025
Device Model Number1012449-15
Device Catalogue Number1012449-15
Device Lot Number20929G1
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 12/09/2022
Supplement Date Received by Manufacturer01/03/2023
Initial Report FDA Received Date12/27/2022
Supplement Report FDA Received Date01/18/2023
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured09/29/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-