| Model Number |
120-440 |
| Medical Device Problem Codes |
Device Appears to Trigger Rejection (1524); Material Rupture (1546)
|
| Health Effect - Clinical Codes |
Capsular Contracture (1761); Failure of Implant (1924)
|
| Date of Event |
12/06/2022
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
Healthcare professional reported left side exchange due to recalled implants and "rupture".Device remains implanted.
|
| |
|
Additional Manufacturer Narrative
|
|
(b)(6).A review of the device history record has been initiated.If any new, changed or corrected information is noted, a supplemental medwatch will be submitted.Further information from the reporter regarding event, product, or patient details has been requested.No additional information is available at this time.Reason for reoperation: rupture, recalled implants.
|
| |
|
Event or Problem Description
|
|
Healthcare professional later reported capsular contracture, baker grade unknown.The device has been explanted.
|
| |
|
Additional Manufacturer Narrative
|
|
The event of capsular contracture is a physiological complication and analysis of the device generally does not assist allergan in determining a probable cause for this event.Reason for reoperation: capsular contracture, baker grade unknown.
|
| |
|
Additional Manufacturer Narrative
|
|
A review of the device history record has been completed.No deviations or non-conformances noted.Device evaluation: based on the device analysis grid, the assessments of the complaint are: rupture: observed, broken on the posterior side assessed as unidentified (tear) opening (shell thickness within specification).Capsular contracture: unable to observe as it is a medical event not related to the device.Anxiety¿product/procedure: unable to observe as it is a medical event not related to the device.Additional observations: creases are observed.Wear abrasion is observed.No further actions are required as the device was implanted.
|
| |
|
Event or Problem Description
|
|
Healthcare professional reported left side exchange due to recalled implants and "rupture".Healthcare professional later reported capsular contracture, baker grade unknown.Device has been explanted.
|
| |
|
Search Alerts/Recalls
|