| Model Number |
MWD156 |
| Medical Device Problem Code |
Unintended System Motion (1430)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
12/05/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Once the investigation has been completed any additional information will be reported in a supplemental report.
|
| |
|
Event or Problem Description
|
|
It was reported that the reaming of the joint fossa was not done properly due to a large shake of the reamer shaft.And whether due to this or simply a bad blade on the 25mm bp reamer, the reaming of the joint fossa was not done properly.This caused difficulty in screw entry and insufficient the base plate placement (contact area) on the joint surface.
|
| |
|
Event or Problem Description
|
|
It was reported that the reaming of the joint fossa was not done properly due to a large shake of the reamer shaft.And whether due to this or simply a bad blade on the 25mm bp reamer, the reaming of the joint fossa was not done properly.This caused difficulty in screw entry and insufficient the base plate placement (contact area) on the joint surface.
|
| |
|
Additional Manufacturer Narrative
|
|
The reported event could not be confirmed, since the returned device is conforming to specifications and fully functional.The device inspection revealed the following: the returned device has evident scratch marks at the distal as well as proximal part.The teeth are slightly bent and dull in appearance due to excessive use of the device.The pre-surgical functional test of the returned device reveals that the components assembled without any hinderance, and it worked well.A review of the device history for the reported lot did not indicate any abnormalities.No corrective actions are required at this time.No indications of material, manufacturing or design related problems were found during the investigation.A review of the labeling did not indicate any abnormalities.Based on investigation, the root cause was attributed to a user related issue.The failure was caused by prolong usage of device.The failure is caused by definitive wear and tear marks which are a result of prolong usage of device over the time involving repetitive cleaning and sterilization cycles.If any further information is provided, the complaint report will be updated.
|
| |
|