APIFIX LTD. MID-C 115; POSTERIOR RATCHETING ROD SYSTEM Back to Search Results
Model Number MID-C 115
Medical Device Problem Code Migration (4003)
Health Effect - Clinical Code Pain (1994)
Date of Event 11/29/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
A review of the device history record confirmed that the device was manufactured and tested according to relevant procedure and shipped according to manufacturer's specifications.On 30-nov-2022 apifix was notified that patient #(b)(6) had a revision surgery performed on (b)(6) 2022 to remove the mid-c following complaints of pain.Patient #(b)(6), index procedure was performed on (b)(6) 2021.Four (4) weeks post-op x-rays demonstrated that the extender kite angle was more than 35° which is not within apifix's described surgical technique."the extender angle should be between 5°-15° degrees compared to the mid-c rod." on (b)(6) 2022, patient #(b)(6) was revised due to device migration resulting in an increase of the kite angle since initial implant.During the revision "the second vertebra screw was removed and upsized to a larger screw.The 1st vertebra screw remained in place.The distal screw remained in place.The original mid-c was explanted and replaced with a new longer mid-c device." on 30-nov-2022, apifix was notified that the patient came into the clinic on (b)(6) 2022 to see the surgeon about back pains.X-rays were taken which showed migration of the second proximal screw.The implant was removed at (b)(6) the following day on (b)(6) 2022.Surgeons (dr.(b)(6) and dr.(b)(6)) had moved hospitals, hence the different facility, (b)(6).Dr.(b)(6) performed the revision as dr.(b)(6) was sick.All instrumentation was removed as the distal poly screw migrated laterally and there was concern about the screw damaging the lung.During the revision it was confirmed that the lung was not damaged.Apifix clinical affairs identified that during the patient's revision (b)(6), the surgeon replaced the screw at t10 with the same size screw and positioned it in the same place; therefore, there may not have been screw purchase which likely led to the screw migration.The standard of care is to place a larger screw into the same place or reposition the implant as the pedicle might be compromised.Reoperation events are a known risk that was assessed and recorded by the product risk assessment dms-777 rev r; this complaint does not change the occurrences rate.The event of screw misplacement/migration is addressed in the ifu (dms-4472 rev g) as potential risks associated with the mid-c system and spinal surgery generally at the time of this report, the incident rate for screw misplacement/migration was 2.80% and is in line with the rate reported in the literature for this type of complication as described in the company's clinical evaluation report (0.2%-13%) (cer dms-727 rev u).
 
Event or Problem Description
On 30-nov-2022 apifix was notified that patient #(b)(6) had a revision surgery performed on (b)(6) 2022 to remove the mid-c following complaints of pain.According to the reporter, x-rays were taken which showed migration of the second proximal screw.
 
Additional Manufacturer Narrative
Return analysis: upon receiving the explanted devices at (b)(6) headquarters ((b)(6), in) the devices were cleaned, sterilized, and evaluated by engineering.Photographs of the device were taken for recording purposes.The [pas] retrieval analysis was not conducted as the device did not meet the criteria of being implanted for more than one year.No defects or damage to the devices were observed.There was no visible wear on the spherical ring of the base.The spherical ring of the pole moved freely.The removal of the device does not appear to be related to the performance of the device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MID-C 115
Common Device Name
POSTERIOR RATCHETING ROD SYSTEM
Manufacturer (Section D)
APIFIX LTD.
1 hacarmel street
kochav yokneam bldg
yokneam elit, 20692 07
IS  2069207
Manufacturer (Section G)
APIFIX LTD.
1 hacarmel street
kochav yokneam bldg
yokneam elit, 20692 07
IS   2069207
Manufacturer Contact
alan vaisman
1 hacarmel st.
kochav yokneam bldg
yokneam ellit, 20692-07
IS   2069207
MDR Report Key16058667
Report Number3013461531-2022-00066
Device Sequence Number17697049
Product Code QGP
UDI-Device Identifier07290018128039
UDI-Public07290018128039
Combination Product (Y/N)N
Initial Reporter StateGA
Initial Reporter CountryUS
PMA/510(K) Number
H170001
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2021
Device Explanted Year2022
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Other
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 07/09/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Model NumberMID-C 115
Device Catalogue NumberMUS-115-040
Device Lot NumberAF-11-046-20
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 11/30/2022
Supplement Date Received by Manufacturer11/30/2022
Initial Report FDA Received Date12/28/2022
Supplement Report FDA Received Date07/09/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured05/04/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient Age19 YR
Patient SexMale
-
-