• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: LIVANOVA DEUTSCHLAND HEATER-COOLER SYSTEM 3T; CONTROLLER, TEMPERATURE, CARDIOPULMONARY BYPASS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

LIVANOVA DEUTSCHLAND HEATER-COOLER SYSTEM 3T; CONTROLLER, TEMPERATURE, CARDIOPULMONARY BYPASS Back to Search Results
Model Number UNKNOWN
Device Problem Microbial Contamination of Device (2303)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 03/14/2022
Event Type  malfunction  
Event Description
On 30 november 2022, livanova has received the medwatch5113446 informing livanova that an heater-cooler system 3t was found contaminated with ntm (nontuberculous mycobacteria).There was no patient involvement.
 
Manufacturer Narrative
There was no known patient involvement.The model, sn and udi of the device have not been provided.Since the sn of the device has not been provided, the manufacturing date is unknown.Livanova (b)(4) implemented a field safety notice for disinfection and cleaning of heater-cooler devices.The z number is z-2076/2081-2015.Livanova (b)(4) manufactures the heater-cooler system 3t.The incident occurred in united states.If any additional information pertinent to the reported event is received, it will be provided in a supplemental report.
 
Event Description
See initial report.
 
Manufacturer Narrative
H10: analysis of all the service activities performed at this customer facility over recent years revealed that three (3) different units were installed: part number 16-02-85, serial numbers (b)(6).Through follow-up communication with a livanova field service representative, it was learned that all the units above mentioned are any longer at customer facility since beginning of 2023.No additional data about this complaint could be retrieved since these units have been out of service for months.Review of complaints database identified no other previous contamination events of heater cooler 3t devices occurred at the involved customer facility.Based on the information available, source of device contamination could not be clearly established and remains unknown.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HEATER-COOLER SYSTEM 3T
Type of Device
CONTROLLER, TEMPERATURE, CARDIOPULMONARY BYPASS
Manufacturer (Section D)
LIVANOVA DEUTSCHLAND
lindberghstrasse 25
munich
Manufacturer (Section G)
LIVANOVA DEUTSCHLAND
lindberghstrasse 25
munich 80939
GM   80939
Manufacturer Contact
enrico greco
14401 w. 65th way
arvada, CO 80004
MDR Report Key16059173
MDR Text Key308400782
Report Number9611109-2022-00692
Device Sequence Number1
Product Code DWC
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K191402
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional
Reporter Occupation Other
Remedial Action Recall
Type of Report Initial,Followup
Report Date 11/30/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/28/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberUNKNOWN
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Date Manufacturer Received09/08/2023
Was Device Evaluated by Manufacturer? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Removal/Correction NumberZ-2076/2081-2015
Patient Sequence Number1
-
-